Emerson Rosemount 0065 Quick Start Manual page 9

Sensor assembly
Hide thumbs Also See for Rosemount 0065:

Advertisement

August 2017
2. The non-metallic label may store an electrostatic charge and become a source of
ignition in Group III environments.
3. Guard the LCD display cover against impact energies greater than 4 joules.
4. Flameproof joints are not intended for repair.
5. A suitable certified Ex d or Ex tb enclosure is required to be connected to temperature
probes with Enclosure option "N".
6. Care shall be taken by the end user to ensure that the external surface temperature on
the equipment and the neck of DIN Style Sensor probe does not exceed 130 °C.
7. Non-Standard Paint options may cause risk from electrostatic discharge. Avoid
installations that cause electrostatic build-up on painted surfaces, and only clean the
painted surfaces with a damp cloth. If paint is ordered through a special option code,
contact the manufacturer for more information.
I1
ATEX Intrinsic Safety
Certificate: Baseefa16ATEX0101X
Standards: EN 60079-0:2012+A11:2013, EN 607960079-11:2012
Markings:
Thermocouples; P
RTDs; P
RTDs; P
Special Condition for Safe Use (X):
1. The equipment must be installed in an enclosure which affords it a degree of ingress
protection of at least IP20.
N1 ATEX Type n
Certificate: BAS00ATEX3145
Standards: EN 60079-0:2012, EN 60079-15:2010
Markings:
ND ATEX Dust
Certificate: FM12ATEX0065X
Standards: EN 60079-0:2012+A11:2013; EN 60079-31: 2014
Markings:
Special Conditions for Safe Use (X):
1. See certificate for ambient temperature range.
2. The non-metallic label may store an electrostatic charge and become a source of
ignition in Group III environments.
3. Guard the LCD display cover against impact energies greater than 4 joules.
4. Flameproof joints are not intended for repair.
5. A suitable certified Ex d or Ex tb enclosure is required to be connected to temperature
probes with Enclosure option "N".
6. Care shall be taken by the end user to ensure that the external surface temperature on
the equipment and the neck of DIN Style Sensor probe does not exceed 130 °C.
7. Non-Standard Paint options may cause risk from electrostatic discharge. Avoid
installations that cause electrostatic build-up on painted surfaces, and only clean the
painted surfaces with a damp cloth. If paint is ordered through a special option code,
contact the manufacturer for more information.
II 1 G Ex ia IIC T5/T6 Ga (see certificate for schedule)
= 500 mW
i
= 192 mW
i
= 290 mW
i
II 3 G Ex nA IIC T5 Gc (–40 °C ≤ T
II 2 D Ex tb IIIC T130 °C Db (–40 °C ≤ T
T6 60 °C ≤ T
≤ +70 °C
a
T6 60 °C ≤ T
≤ +70 °C
a
T6 60 °C ≤ T
≤ +60 °C
a
T5 60 °C ≤ T
≤ +70 °C
a
≤ +70 °C)
a
≤ +70 °C)
a