Download Print this page

Wiring Diagrams - Emerson Rosemount 0065 Quick Start Manual

Sensor assembly
Hide thumbs Also See for Rosemount 0065:

Advertisement

August 2017
1.0

Wiring diagrams

Figure 1. Rosemount Series 65 RTD Lead Wire Configuration
Flying leads and spring-loaded adapter (termination codes 0, 1, or 3 only)
Red
Red
White
White
Note
For 3-wire systems use one white and two red leads. Do not connect the white leads.
Insulate or terminate the unused white lead in a manner that prevents shorting to the
ground. For 2-wire systems, connect both sets of leads.
Single element
White
White
Red
Red
Terminal block (termination code 2 and 4)
Single element
4
3
1
1
3
6
6
4
Dual element
Red
Red
Black
Blue
Blue
Green
Dual element
Red
3
Red
2
1
W
2
1
3
Red
6
6
4
5
Red
5
White
4
Quick Start Guide
3

Advertisement

loading

This manual is also suitable for:

Rosemount 0185