Emerson Rosemount 3300HT Instruction Manual

Emerson Rosemount 3300HT Instruction Manual

Perph-x high performance ph/orp sensors
Hide thumbs Also See for Rosemount 3300HT:

Advertisement

Instruction Manual
LIQ-MAN-PERpH-X
Rev. J
April 2017
Rosemount
3300HT/3300HTVP/3400HT/
3400HTVP/3500P/3500VP
PERpH-X High Performance pH/ORP Sensors

Advertisement

Table of Contents
loading

Summary of Contents for Emerson Rosemount 3300HT

  • Page 1 Instruction Manual LIQ-MAN-PERpH-X Rev. J April 2017 Rosemount 3300HT/3300HTVP/3400HT/ ™ 3400HTVP/3500P/3500VP PERpH-X High Performance pH/ORP Sensors...
  • Page 2 hasgkas...
  • Page 3 Essential Instructions Read this page before proceeding! Emerson designs, manufactures and tests its products to meet many national and international stan- dards. Because these sensors are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount products.
  • Page 4 About This Document This manual contains instructions for installation and operation of the Rosemount 3300HT/3300HTVP/ 3400HT/3400HTVP/3500P/3500VP. The following list provides concerning all revisions of this document. Rev. Level Date Notes 04/2017 Updated information with new Emerson Style Guidelines, Updated Ordering Information, Specifications, and Wiring Diagrams. Added Accessories Information, EC Declaration of Conformity and FM Installation Drawings.
  • Page 5: Table Of Contents

    Instruction Manual Table of Contents LIQ-MAN-PERpH-X April 2017 Contents Section 1: Specifications Specifications ......................1 Product Certifications ..................3 Ordering Information...................4 Section 2: Installation Storage........................7 Installation......................7 Retraction with Kit P/N 23240-00 ..............16 Electrical Installation..................17 Section 3: Startup and Calibration Electrode Preparation ..................21 pH Sensor Calibration ..................21 Recommended pH Sensor Standardization ............21 Section 4: Maintenance...
  • Page 6 Table of Contents Instruction Manual April 2017 LIQ-MAN-PERpH-X Table of Contents...
  • Page 7: Section 1: Specifications

    Pt-100 RTD Process Connections Must use 1 in. compression process connector (P/N 23166-00 or 23166-01) Weight/Shipping Weight 1 lb/2 lb (0.5 kg/0.9 kg) Cable Length 15 ft. integral cable (Rosemount 3300HT) or VP8 Cable for Rosemount 3300HTVP (sold separately) Specifications...
  • Page 8 Specifications Instruction Manual April 2017 LIQ-MAN-PERpH-X Table 1-3: Rosemount 3400HT/3400HTVP sensor specifications Measured Range pH range 0 to 14 pH ORP range -1500 mV to 1500 mV Operating Temperature Without Preamplifier 41 to 311 °F (5 to 155 °C) With Preamplifier up to 293 °F (145 °C) Storage Temperature 14 to 138 °F (-10 to 70 °C)
  • Page 9: Product Certifications

    Instruction Manual Specifications LIQ-MAN-PERpH-X April 2017 Product Certifications Please see online certificates for further details. IECEx Sensors without preamp – Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60°C) Sensors with SMART preamp (pH only) – Ex ia IIC T4 Ga (-20°C ≤ Ta ≤ +60°C) Sensors with standard preamp (ORP only) –...
  • Page 10: Ordering Information

    Specifications Instruction Manual April 2017 LIQ-MAN-PERpH-X Ordering Information Table 1-5: Rosemount 3300HT sensor ordering information Model Sensor type 3300HT pH/ORP Sensor Measuring Electrode pH - GPHT Glass O-ring Material EPDM Viton Kalrez Typical Model Number: 3300HT-10-30 Table 1-6: Rosemount 3300HTVP sensor ordering information...
  • Page 11 Instruction Manual Specifications LIQ-MAN-PERpH-X April 2017 Table 1-7: Rosemount 3400HT sensor ordering information Model Sensor type 3300HTVP pH/ORP Sensor Measuring Electrode pH - GPHT Glass Sensor Length 21 in. Titanium Tube 36 in. Titanium Tube O-ring Material EPDM Viton Kalrez Cable Length 9.5 in.
  • Page 12 Specifications Instruction Manual April 2017 LIQ-MAN-PERpH-X Table 1-7: Rosemount 3500P sensor ordering information Model Sensor type 3500P pH/ORP Sensor Electrolyte Selection Bio-Film Resistant High Temperature Metal Resistant Oil Resistant Poisoning Resistant Scaling Resistant Preamplifier/Cable Preamplifier with 25 ft. No Preamplifier with 15 ft. Cable Measuring Electrode Type pH –...
  • Page 13: Section 2: Installation

    It is recommended that electrodes be stored in their original shipping containers until needed. Do not store Rosemount 3500P at temperatures below 14 °F (-10 °C). Rosemount 3300HT, 3300HTVP, 3400HT, and 3400HTVP below 23 °F (-5 °C). Electrodes should be stored with a protective cap containing KCl solution.
  • Page 14 Installation Instruction Manual April 2016 LIQ-MAN-PERpH-X Figure 2-2: Rosemount 3300HT and 3300HTVP sensor dimensional drawing Installation...
  • Page 15 Instruction Manual Installation LIQ-MAN-PERpH-X April 2017 Figure 2-3: Rosemount 3400HT and 3400HTVP sensor dimensions with optional Ball Valve PN 23240-00 Installation...
  • Page 16 Installation Instruction Manual April 2017 LIQ-MAN-PERpH-X Figure 2-4: Rosemount 3400HT and 3400HTVP sensor dimensions with optional Ball Valve PN 23765-00 Calibration and Maintenance...
  • Page 17 Installation LIQ-MAN-PERpH-X April 2017 Figure 2-5: Rosemount 3300HT and 3300HTVP flow through and insertion installation. 1½ in. pipe Tee (PN 2002011) with 1 in. threaded connections. Note: The sensor must be mounted at least 10° above the horizon. Figure 2-6: Rosemount 3500P and 3500VP sensor dimensional drawing...
  • Page 18 Installation Instruction Manual April 2017 LIQ-MAN-PERpH-X Figure 2-7: Rosemount 3500P AND 3500VP Jet Spray Cleaner (PN 12707-00) for Submersion Installation. This accessory is especially useful for keeping the sensor clean in dirty ponds or tanks. It can be mounted using the Handrail Mounting Assembly or a similar submersion accessory. Figure 2-8: Rosemount 3500P and 3500VP flow through installation Installation...
  • Page 19 Instruction Manual Installation LIQ-MAN-PERpH-X April 2017 Figure 2-9: Rosemount 3500P or 3500VP for Submersion Installation using the Handrail Mounting Assembly (PN 11275-01) All parts shown are supplied. Sensor sold separately. Installation...
  • Page 20 Installation Instruction Manual April 2017 LIQ-MAN-PERpH-X Figure 2-10: Rosemount 3400HT and 3400HTVP mounting details – retraction version Figure 2-11: Exploded view of 1½ in. ball valve kit PN 23240-00 used with process connector PN 23166-00 (or PN 23166-01). Ball valve kit includes 11/2" x 1" reducer, 11/2" close nipple, and 11/2" ball valve. Figure 2-12: Exploded view of 1¼...
  • Page 21 Instruction Manual Installation LIQ-MAN-PERpH-X April 2017 Figure 2-13: Rosemount 3300HT or 3300HTVP typical submersion installation Installation...
  • Page 22: Retraction With Kit P/N 23240-00

    Installation Instruction Manual April 2017 LIQ-MAN-PERpH-X Retraction with Kit P/N 23240-00 WARNING System pressure may cause the sensor to blow out with great force unless care is taken during removal. 2.3.1 Rosemount 3400HT/3400HTVP-21 (21 in. Tube) Be certain system pressure at the sensor is below 64 psig (542 kPa) before proceeding with the retraction.
  • Page 23: Electrical Installation

    Instruction Manual Installation LIQ-MAN-PERpH-X April 2017 Electrical Installation For additional wiring information on this product, including sensor combinations not shown here, please refer to Transmitter Wiring Diagrams. Figure 2-14: PERpH-X sensors, preamplifier in instrument Installation...
  • Page 24 Installation Instruction Manual April 2017 LIQ-MAN-PERpH-X Figure 2-15: PERpH-X sensors, preamplifier in junction box Installation...
  • Page 25 Instruction Manual Installation LIQ-MAN-PERpH-X April 2017 Figure 2-16: PERpH-X sensors, preamplifier in sensor Installation...
  • Page 26 Installation Instruction Manual April 2017 LIQ-MAN-PERpH-X Figure 2-17: PERpH-X sensors, preamplifier in sensor – junction box used for extending cable Installation...
  • Page 27: Section 3: Startup And Calibration

    Instruction Manual Startup and Calibration LIQ-MAN-PERpH-X April 2017 Section 3: Startup and Calibration Electrode Preparation Remove electrode from shipping container. Remove the protective boot covering the electrode bulb. Rinse away salt film with clean water, then gently shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there.
  • Page 28 Starup and Calibration Instruction Manual April 2017 LIQ-MAN-PERpH-X Startup and Calibration...
  • Page 29: Section 4: Maintenance

    Instruction Manual Maintenance LIQ-MAN-PERpH-X April 2017 Section 4: Maintenance Electrode Cleaning Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement. To remove oil deposit, clean the electrode with a mild non-abrasive detergent. To remove scale deposits, soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution.
  • Page 30 Maintenance Instruction Manual April 2017 LIQ-MAN-PERpH-X Maintenance...
  • Page 31: Section 5: Accessories

    Instruction Manual Accessories LIQ-MAN-PERpH-X April 2017 Section 5: Accessories Accessories Table 5-1: Rosemount 3300HT/3300HTVP/3400HT/3400HTVP/3500P/3500PVP sensors accessories Part Number Description Rosemount 3300HT/3300HTVP/3400HT/3400HTVP Sensors 23166-00 1 in. MNPT process connector, stainless steel with EPDM O-ring 23166-01 1 in. MNPT process connector, titanium with EPDM O-ring 23796-00 316 SST retraction kit for use with a 1-¼...
  • Page 32 Accessories Instruction Manual April 2017 LIQ-MAN-PERpH-X Part Number Description Solution Kits 24231-00 High temperature solution kit (0 to 145 °C) with EPDM O-rings 24231-01 Bio-film resistant solution kit (0 to 60 °C) with EPDM O-rings 24231-02 Poisoning resistant solution kit (0 to 100 °C) with viton O-rings 24231-03 Oil resistant solution kit (0 to 100 °C) with viton O-rings 24231-04...
  • Page 33: Ec Declaration Of Conformity

    Instruction Manual EC Declaration of Conformity LIQ-MAN-PERpH-X April 2017 EC Declaration of Conformity Note: Please see website for most recent Declaration. EC Declaration of Conformity...
  • Page 34 EC Declaration of Conformity Instruction Manual April 2017 LIQ-MAN-PERpH-X EC Declaration of Conformity...
  • Page 35: Intrisicallly Safe Sensor Installation Drawing - Fm

    Instruction Manual FM Installation LIQ-MAN-PERpH-X April 2017 Intrisicallly Safe Sensor Installation Drawing - FM FM Installation...
  • Page 36 ©2017 Emerson. All rights reserved. Emerson The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of 8200 Market Blvd. one of the Emerson family of companies. All other marks are the property of their respective Chanhassen, MN 55317, owners.