Pipe clamp sensor

Advertisement

Quick Start Guide
Figure 8. Pipe Clamp Sensor Assembly with Protection Inlay
A. Extension of pipe clamp
B. Pipe clamp
C. Sensor
3.2
Replacement sensor
The spring loaded sensor can be ordered for replacement using the Rosemount
0085 Pipe Clamp Sensor
Use the following steps to replace the sensor.
1. Loosen and remove the original sensor from the extension of the pipe clamp.
2. Add pipe compound or PTFE tape (where local piping codes allow) to the
threads of the new sensor.
3. Insert the new sensor into the extension of the pipe clamp sensor and ensure
the sensor tip passes through the hole of the pipe clamp. Refer to
more information.
4. Screw in the sensor and tighten to 24 ft-lbs of torque.
8
A
C
D
D. Corrosion protection inlay
E. Pipe
Product Data
Sheet.
B
E

August 2017

Figure 6
for