Series 2280 Precision Measurement DC Power Supplies Reference Manual

:STATus:PRESet

This command resets all the device-dependent enable registers.
Type
Command only
Usage
:STATus:PRESet
Details
Sending this command can reset the bits of the following enable registers to 0:
Operation Event Enable Register
Operation Instrument Event Enable Register
Operation Instrument Summary Event Enable Register
Measurement Event Enable Register
Measurement Instrument Event Enable Register
Measurement Instrument Summary Event Enable Register
Questionable Event Enable Register
Questionable Instrument Event Enable Register
Questionable Instrument Summary Event Enable Register
Example
:STAT:PRES
Also see
None

:STATus:QUEStionable[:EVENt]?

This command reads the Questionable Event Register.
Type
Query only
Usage
:STATus:QUEStionable[:EVENt]?
Details
This command reads the Questionable Event Register of the status model and then clears the
register.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register. For details, see
Example
:STAT:QUES?
077085501 / February 2015
Affected by
Not applicable
Affected by
Not applicable
Questionable Event Register
Returns the contents of Questionable Event Register
Section 7: SCPI command reference
Where saved
Not applicable
Reset all the enable registers to 0.
Where saved
Not applicable
(on page D-12).
Default value
Not applicable
Default value
Not applicable