Series 2280 Precision Measurement DC Power Supplies Reference Manual
Example
:MEAS:CURR? 10,5
Also see
:CONFigure[n]:<function>

:READ[n]?

:READ[n]?
This query makes measurements and returns the first reading.
Type
Query only
Usage
:READ[n]?
<n>
Details
When this command is sent, the following commands execute in the order they are presented:
:ABORt
:INITiate
:FETCh?
When this command is executed, the instrument resets the trigger model. If the instrument is in the
idle state, the :INITiate command takes the instrument out of the idle state. The :FETCh?
command reads the first post-processed reading.
Also see
:ABORt[n]
:FETCh[n]?
:INITiate[n][:IMMediate]

ARM subsystem

The commands in this subsystem configure and control the trigger model for the arm layer.
077085501 / February 2015
(on page 7-11)
(on page 7-17)
Affected by
Not applicable
Channel number: For single-channel instruments, the value is 1.
(on page 7-11)
(on page 7-12)
(on page 7-45)
Measure current on the10 A range with 5.5 digit display
resolution.
Example output:
2.473375E-02A