Section 7: SCPI command reference

*CLS

This command clears the event registers and error queues.
Type
Command only
Usage
*CLS
Details
Use *CLS to clear the event registers and queues. It performs the following actions:
Clear the Standard Event Status Register and Status Byte Register.
Clear the event register of Questionable Event, Questionable Instrument Event and Questionable
Instrument Summary Event.
Clear the event register of Operation Event, Operation Instrument Event and Operation Instrument
Summary Event.
Clear the event register of Measurement Event, Measurement Instrument Event and Measurement
Instrument Summary Event.
Clear the error queue.
To reset all of the bits of the Standard Event Enable Register, send the command:

*ESE 0

Also see
*ESE
(on page 7-2)
STATus subsystem
*ESE
This command sets and queries bits in the Standard Event Status Enable Register.
Type
Command and query
Usage
*ESE <n>
*ESE?
<n>
7-2
Series 2280 Precision Measurement DC Power Supplies Reference Manual
Affected by
Not applicable
(on page 7-82)
Affected by
Power cycle
The value of the Standard Event Status Enable Register (0 to 255)
Where saved
Not applicable
Where saved
Not saved
Default value
Not applicable
Default value
See Details