Section 7: SCPI command reference

:TRIGger[:SEQuence[n]]:SAMPle:COUNt

This command sets the number of measurement per trigger event.
Type
Command and query
Usage
:TRIGger[:SEQuence[n]]:SAMPle:COUNt <count>
:TRIGger[:SEQuence[n]]:SAMPle:COUNt?
:TRIGger[:SEQuence[n]]:SAMPle:COUNt? DEFault
:TRIGger[:SEQuence[n]]:SAMPle:COUNt? MINimum
:TRIGger[:SEQuence[n]]:SAMPle:COUNt? MAXimum
[n]
<count>
Details
This command sets the number of measurement per trigger event. For details, refer to
(on page 3-21).
Example
:TRIG:SAMP:COUN 10
Also see
Trigger model

:TRIGger[:SEQuence[n]]:SOURce

This command selects the control source for trigger layer.
Type
Command and query
Usage
:TRIGger[:SEQuence[n]]:SOURce <source>
:TRIGger[:SEQuence[n]]:SOURce?
[n]
<source>
7-124
Series 2280 Precision Measurement DC Power Supplies Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1.
The number of measurements: 1 to 2500.
(on page 3-21)
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1.
The trigger source for trigger layer:
Immediate triggering: IMMediate. No delay or gating in the
detection layer
External triggering: EXTernal.
Manual triggering: MANual. The trigger signal is generated by
pressing the TRIGGER key
Where saved
Save settings
Set the sample count to 10.
Where saved
Save settings
Default value
1
Trigger model
Default value
IMM
077085501 / February 2015