Series 2280 Precision Measurement DC Power Supplies Reference Manual
Detail
This command allows you to save the present instrument setup to the external USB flash drive.
Make sure that the external USB flash drive is connected to the instrument before you send this
command. An unstable connection between the USB flash drive and the instrument will cause an
error 520, "No USB flash drive found."
Example
:MMEMory:SAVE:SETup "mysetup"
Also see
:MMEMory:LOAD:SETup
*RCL
(on page 7-5)
*SAV
(on page 7-6)
OUTPut subsystem
The output subsystem provides information and settings that control the source output.
:OUTPut:DELay:FALLing
This command specifies a delay that the instrument waits before turning off the specified output.
Type
Command and query
Usage
:OUTPut:DELay:FALLing <n>
:OUTPut:DELay:FALLing <n>,<channel>
:OUTPut:DELay:FALLing?
:OUTPut:DELay:FALLing? <channel>
:OUTPut:DELay:FALLing? DEFault
:OUTPut:DELay:FALLing? DEFault,<channel>
:OUTPut:DELay:FALLing? MINimum
:OUTPut:DELay:FALLing? MINimum,<channel>
:OUTPut:DELay:FALLing? MAXimum
:OUTPut:DELay:FALLing? MAXimum,<channel>
<n>
<channel>
077085501 / February 2015
(on page 7-46)
Affected by
Recall settings
Instrument reset
Power cycle
Delay rising time: 0 to 10 s with the resolution of 0.001 s.
The channel number:
Channel 1: CH1
Channel 2: CH2
The default value is CH1
Section 7: SCPI command reference
Save the present setup to the external USB flash
drive as "mysetup".
Where saved
Save settings
Default value
0
7-47