Sense[N]:<Function>:Average:count - Keithley 2280 Series Reference Manual

Precision measurement
Hide thumbs Also See for 2280 Series:
Table of Contents

Advertisement

Section 7: SCPI command reference
Usage
:SENSe[n]:FUNCtion <name>
:SENSe[n]:FUNCtion?
[n]
<name>
Details
Set this command to the type of measurement you want to make.
Reading this command returns the measure function that is presently active.
Examples
:SENS:FUNC "VOLT"
Also see
None
:SENSe[n]:<function>:AVERage:COUNt
This command sets the number of measurements that are averaged when filtering is enabled for the specified
function.
Type
Command and query
Usage
:SENSe[n]:<function>:AVERage:COUNt <value>
:SENSe[n]:<function>:AVERage:COUNt?
:SENSe[n]:<function>:AVERage:COUNt? DEFault
:SENSe[n]:<function>:AVERage:COUNt? MINimum
:SENSe[n]:<function>:AVERage:COUNt? MAXimum
[n]
<function>
<value>
7-52
Series 2280 Precision Measurement DC Power Supplies Reference Manual
Channel number: For single-channel instruments, the value is 1.
A string that contains the measurement function to make active:
Concurrent: "CONCurrent[:DC]"
Current: "CURRent[:DC]"
Voltage: "VOLTage[:DC]"
For details, see
Select a measurement function
Affected by
Recall settings
Instrument reset
Power cycle
Channel number: For single-channel instruments, the value is 1.
The measurement function to which this setting applies:
Concurrent: CONCurrent[:DC]
Current: CURRent[:DC]
Voltage: VOLTage[:DC]
The number of readings required for each filtered measurement (2 to
100)
(on page 2-65).
Make the voltage measurement function the active
function.
Where saved
Save settings
Default value
10
077085501 / February 2015