Series 2280 Precision Measurement DC Power Supplies Reference Manual
:STATus:MEASurement[:EVENt]?
This command reads and clears the Measurement Event Register of the status model.
Type
Query only
Usage
:STATus:MEASurement[:EVENt]?
Details
This attribute reads the Measurement Event Register of the status model and then clears the register.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register. For details, see
Also see
None
:STATus:MEASurement:ENABle
This command sets or reads the contents of the Measurement Event Enable Register of the status model.
Type
Command and query
Usage
:STATus:MEASurement:ENABle <n>
:STATus:MEASurement:ENABle?
<n>
Details
This command sets or reads the contents of the enable register of the Measurement Event Register.
When one of these bits is set and the corresponding bit in the Measurement Event Register is set, the
MSB bit in the Status Byte Register is set.
When sending binary values, preface <n> with #b. When sending hexadecimal values, preface <n>
with #h. No preface is needed when sending decimal values.
Example
:STAT:MEAS:ENAB #b0010000000000000
Also see
Measurement Event Register
077085501 / February 2015
Affected by
Not applicable
Measurement Event Register
Affected by
Power cycle
The status of the Measurement Status Register
(on page D-15)
Section 7: SCPI command reference
Where saved
Not applicable
(on page D-15).
Where saved
Not applicable
Sets the ISUM bit of the measurement
status enable register using a binary
value.
You could also send the decimal value
8192 or the hexadecimal value: #h2000.
Default value
Not applicable
Default value
0