Non-incendive field wire parameters of vortex flowmeter (DY)
and vortex flow converter (DYA).
Ui (Vmax)=30
Ii (Imax)=165mA
Ci=12nF
Li=0.15mH
Installation requirement between flowmeter, converter and
general purpose equirement.
Uo
Ui
Io
Ii
Po
Lo
Li+Lcable
Voc
Vmax
Isc
Imax
La
Li+Lcable
Uo, Io, Po, Co Voc, Isc, Ca and La are nonincendive field
wire parameters of general purpose equipment.
WARNING
• The general purpose equipment must be CSA
certified as the equipment which have type n or
non-incendive field wire parameters.
• Installation should be in accordance with
Canadian Electrical Code Part I.
• Dust-tight conduit seal must be used when
installed in class II and III environments.
• Do not alter drawing without authorization from
CSA.
Non Hazardous
Location
General Purpose
Equipment
+
+
Power
Supply
SUPPLY
PULSE
+
+
Receiver
+
+
Power
Supply
SUPPLY
PULSE
+
+
Receiver
: installed
Pi (Pmax)=0.9W
Pi
Co
Ci+Ccable
Ca
Ci+Ccable
F100401_2.EPS
10. EXPLOSION PROTECTED TYPE INSTRUMENT

10.4.6 Dual Seal (Option: /CF11, /CS11)

Dual Seal:
Certified by CSA to the requirement of ANSI/ISA 12.27.01
No additional sealing required.
Primary seal failure annunciation: at the O-ring seal portion
between shedder bar and amplifier housing.
+
+
+
+
10-11
IM 01F06A00-01E