English - GE Autosampler Quick Start Manual

For use with sievers m-series laboratory and portable analyzers
Hide thumbs Also See for Autosampler:

Advertisement

ENGLISH

WARNINGS
For continued protection against fire hazard, replace fuse with
same type and rating.
This is a Safety Class 1 product. It must be attached to a grounded
power source. If this instrument is used in a manner not specified by GE
Analytical Instruments USA, the protection provided by the instrument
may be impaired.
Always stop TOC measurements/protocols before turning off or
unplugging the GE Autosampler.
Observe the following guidelines when using the GE Autosampler
in conjunction with flammable liquids. To reduce the risk of fire, the
GE Autosampler shall not be used to sample flammable liquids with a
flash point of less than ambient temperature + 25 ºC. Other than the
needle, the GE Autosampler should have no parts in contact with the
fluid. Ensure that no parts, in or near the fluid surface are capable of
igniting a flammable liquid.
Keep clear of the probe arm during operation. Although it is designed to
stop upon detecting an obstruction, the probe arm can move without
warning during normal operation and at start-up.
This symbol on the instrument indicates that the product does not
contain any restricted substances included in China RoHS I.
PORTUGUÊS
AVISOS
Substitua os fusíveis pelo mesmo tipo e categoria para uma pro
teção constante contra risco de incêndio.
Este produto está classificado com Segurança Classe 1. Ele deve estar
conectado a uma fonte de energia aterrada. Se este instrumento for
usado de forma diferente da especificada pela GE Analytical Instruments
USA, a proteção fornecida pelo instrumento será prejudicada.
Sempre pare os protocolos/medidas de TOC antes de desligar e
desconectar o amostrador automático GE.
Siga as orientações a seguir quando usar o amostrador automático GE
com líquidos inflamáveis. Para reduzir o risco de incêndios, o amostrador
automático GE não deve ser usado para amostrar líquidos inflamáveis
com ponto de inflamação menor do que a temperatura ambiente de
+ 25 ºC. Com exceção da agulha, o amostrador automático GE não
deve ter qualquer parte em contato com o fluido. Certifique-se de
que nenhuma parte, na superfície ou próxima ao fluido, seja capaz de
incendiar um líquido inflamável.
Mantenha-se afastado do braço da sonda durante a operação. Embora
tenha sido projetado para parar quando detectar uma obstrução, o
braço da sonda pode se mover sem aviso durante a operação normal
e na partida.
20
2016 General Electric Company. All rights reserved.
3.
Position the needle into the arm, lowering the tip through the
lower mount block, and pushing the top of the assembly up-
ward and into the upper mount block.
4.
Secure with the nut (placed over the washer), and tighten with
1/2" open-end wrench. Be careful NOT to over-tighten the nut.
5.
Screw the PEEK nut (with sample tubing attached) back into the
threaded mount of the needle assembly until it is finger tight.
DO NOT over-tighten the nut.
6.
Using a flat-blade screwdriver, raise the arm back into place.
STEP 5: Route the Sample Tubing
1.
Insert the anchor guide into the hole on the top side of the
GE Autosampler.
Anchor Guide
2.
Route the tubing through the small eyelet on the Autosampler
arm.
Small Eyelet on Needle Arm
3.
Connect the Valco fitting to the end of the tubing.
4.
Route the tubing through the anchor guide.
DQS 77700-01 MUL Rev. B
GE Autosampler (with M-Series Analyzers Quick Start Guide)
STEP 6: Connect the Communication/Power Cables
Network
Ethernet
Ethernet
USB-USB
[Straight-Through
Cable]
Computer
Sievers Analyzer GE Autosampler
Connecting Directly to a Network
USB-USB
USB-USB
Printer
Computer
Analyzer
Autosampler
Connecting to a Computer
1.
Connect the GE Autosampler to the Analyzer using the
provided USB-to-USB cable.
2.
Connect the Analyzer to a network using a straight-thru
Ethernet
cable or directly to a computer using a USB
cable.
Insert
the
power
cord
into
the
Autosampler and then connect to a grounded power outlet.
STEP 7: Install DataPro2 Software
Refer to the GE Autosampler Installation Guide for instructions.
(Download from www.geinstruments.com/library/manuals).
www.geinstruments.com
GE
5