Do not use the ALTAIR 5X PID Multigas Detector to test for combustible gases in atmospheres
containing vapors from liquids with a high flash point (above 38 °C, 100 °F) as this may result in
erroneously low readings.
-
Allow sufficient time for device to display accurate reading. Response times vary based on the
type of sensor being utilized (→ chapter 6.2). Allow a minimum of 1 second per foot (3 seconds
per meter) of sample line to allow the sample to be drawn through the sensors.
-
Sampling lines made from 0.062 inch (1.57 mm) inner diameter tubing provide fast transport times
to the device; however, they must be limited to 50 feet (15 m) in length.
-
Sampling of reactive toxic gases (Cl
line and probe kits listed in chapter 9.
-
All device readings and information must be interpreted by someone trained and qualified in inter-
preting device readings in relation to the specific environment, industrial practice and exposure
limitations.
Observe Proper Battery Maintenance
Use only battery chargers made available by MSA for use with this device; other chargers may damage
the battery pack and the device. Dispose of in accordance with local health and safety regulations.
Be Aware of Environmental Conditions
A number of environmental factors may affect the sensor readings, including changes in pressure,
humidity and temperature. Pressure and humidity changes also affect the amount of oxygen actually
present in the atmosphere.
Be Aware of the Procedures for Handling Electrostatically Sensitive Electronics
The device contains electrostatically sensitive components. Do not open or repair the device without
using appropriate electrostatic discharge (ESD) protection. The warranty does not cover damage
caused by electrostatic discharges.
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
This device complies with part 15 of the FCC Rules. Operation is subject to the following
conditions: (1) This device may not cause harmful interference, and (2) this device must
accept any interference received, including any interference that may cause undesired
operation.
This is a class A product in accordance with CISPR 22. In a domestic environment, this product may
cause radio interference, in which case the user may be required to take adequate measures.
This Class A digital apparatus complies with Canadian ICES-003.
Be Aware of the Warranty Regulations
The warranties made by Mine Safety Appliances Company with respect to the product are voided if the
product is not used and maintained in accordance with the instructions in this manual. Please protect
yourself and others by following them. We encourage our customers to write or call regarding this
equipment prior to use or for any additional information relative to use or service.
Be Aware of the Product Regulations
Follow all relevant national regulations applicable in the country of use.
, ClO
, NH
) must only be done with the reactive gas sample
2
2
3
WARNING
ALTAIR 5X PID
Safety Regulations
US
9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents