Safety Regulations

1.1

Correct Use

The ALTAIR 5X PID Multigas Detector, hereafter also referred to as device, is for use by trained and
qualified personnel. The device is designed to be used when performing a hazard assessment to:
-
Assess potential worker exposure to combustible and toxic gases and vapors as well as low level
of oxygen.
-
Determine the appropriate gas and vapor monitoring needed for a workplace.
The ALTAIR 5X PID Multigas Detector can be equipped to detect:
-
Combustible gases and certain combustible vapors.
-
Volatile organic compounds (VOC).
-
Oxygen-deficient or oxygen-rich atmospheres.
-
Specific toxic gases for which a sensor is installed.
-
While the device can detect up to 30 % oxygen in ambient air, it is approved for use only up to
21 % oxygen.
-
Perform a blocked flow test before each day's use.
-
It is recommended that a Bump Test is performed before each day's use; adjust if necessary.
-
Perform a Bump Test more frequently if exposed to silicone, silicates, lead-containing
compounds, hydrogen sulfide, or high contaminant levels.
-
Recheck calibration if unit is subjected to physical shock.
-
Use only to detect gases/vapors for which a sensor is installed.
-
Do not use to detect combustible dusts or mists.
-
For accurate catalytic combustible readings, make sure adequate oxygen is present
(>10 % O
).
2
-
Never block pump inlet, except to perform a sampling system safety test.
-
Have a trained and qualified person interpret device readings.
-
Risk of Explosion: Do not remove battery pack or recharge Li Ion battery in a hazardous location.
-
Do not alter or modify device.
-
Use only MSA-approved sampling lines.
-
Do not use silicone tubing or sampling lines.
-
Wait sufficient time for the reading; response times vary based on gas and length of sampling
line.
-
Properly identify the VOC gas being measured before using VOC response factors or setting
alarm values (exposures, STEL, TWA)
-
Recognize that the VOC display readings are in increments of 0.1ppm from 0-999 ppm, then
1 ppm increments from 1000-2000 ppm with a Response Factor of one.
-
Ensure installed PID lamp corresponds to the PID lamp setting on the display shown at startup.
Incorrect use can cause death or serious personal injury.
It is imperative that this operating manual be read and observed when using the product. In particular,
the safety instructions, as well as the information for the use and operation of the product, must be
carefully read and observed. Furthermore, the national regulations applicable in the user's country
must be taken into account for safe use.
Alternative use, or use outside this specification will be considered as non-compliance. This also
applies especially to unauthorized alterations to the product and to commissioning work that has not
been carried out by MSA or authorized persons.
WARNING
ALTAIR 5X PID
Safety Regulations