Bump Test (BUMP page)
This page allows the user to perform an automated Bump Test on the device. To perform the test, the
(YES) button is pressed. See chapter 4.9 for details on performing the Bump Test.
If the
button is pressed, the Bump Test is not performed, and the display shows the next page in
the sequence (PEAK).
If the
button is pressed, the Bump Test is not performed, and the display reverts back to the normal
MEASURE page.
Peak Readings (PEAK page)
This page shows the highest levels of gas recorded by the device since turn-ON or since peak readings
were reset.
To reset the peak readings:
(1)
Access the PEAK page.
(2)
Press the
button.
This page can be de-activated through MSA Link software.
Minimum Readings (MIN page)
This page shows the lowest level of oxygen recorded by the device since turn-ON or since the MIN
reading was reset. It is only shown if an oxygen sensor is installed and enabled.
To reset the MIN reading:
(1)
Access the MIN page.
(2)
Press the
button.
Short Term Exposure Limits (STEL page)
If the STEL alarm activates, leave the contaminated area immediately; the ambient gas concentra-
tion has reached the preset STEL alarm level. Failure to follow this warning will cause over-exposure
to toxic gases and persons relying on this product for their safety could sustain serious personal
injury or death.
This page shows the average exposure over a 15-minute period.
When the amount of gas detected by device is greater than the STEL limit:
-
Alarm sounds, alarm lights flash.
-
Alarm LEDs flash
-
"STEL ALARM" message flashes.
To reset the STEL:
(1)
Access the STEL page.
(2)
Press the
button.
The STEL alarm is calculated over a 15-minute exposure.
STEL calculation examples:
Assume the device has been running for at least 15 minutes:
WARNING
ALTAIR 5X PID
Description
US
21