Setting MotionAlert Access
Setting this parameter allows access to the MOTIONALERT page from the MEASURE page.
If access is denied here:
-
the user cannot access the MOTIONALERT page to enable or disable that feature
-
the InstantAlert feature (chapter 3.3) cannot be activated.
(1)
To grant or deny user access to the MOTIONALERT page, use the button to change the indi-
cated selection.
User access is:
permitted when the setting indicates ON.
denied when the setting indicates OFF.
(2)
The selection is confirmed by pressing either the
Setting Sensor Alarms
This page allows modifying the preset alarm values of:
-
LOW Alarm
-
HIGH Alarm
-
STEL Alarm
-
TWA Alarm.
Factory set alarm levels are shown in chapter 6.1.
(1)
Press the q button to enter Sensor Alarm setup.
LOW Alarm Setup screen displays.
Fig. 8
Sensor Alarm Setup
(2)
Press
the
button to abort the operation or
the
button to go to next alarm setup or
the q button to change the alarm setpoints.
Alarm Value for the first Sensor displays.
Fig. 9
Sensor Alarm Setup
or
button.
ALTAIR 5X PID
Operation
US
36