Low Battery Warning
If battery warning alarm activates while using the device, leave the area immediately as the end of
battery life is approaching. Failure to follow this warning can result in serious personal injury or death.
Fig. 4
Battery Warning
The duration of remaining device operation during a Low Battery Warning depends on ambient
temperatures, battery condition alarm status. Nominal battery life is 30-60 minutes after the Battery
Warning activates.
When the device goes into battery warning the:
-
battery life indicator continuously blinks
-
alarm sounds and alarm LEDs flash every 30 seconds
-
Safe LED no longer flashes
-
device continues to operate until it is turned OFF or battery shutdown occurs.
Battery Shut Down
If battery alarm displays, stop using the device as it no longer has enough power to indicate potential
hazards, and persons relying on this product for their safety could sustain serious personal injury or
death.
The device goes into battery shutdown mode 60 seconds before final shutdown (when the batteries
can no longer operate the device):
-
"BATTERY ALARM" flashes on the display
-
Alarm sounds
-
Alarm LEDs flash
-
Fault LED is on
-
No other pages can be viewed; after approximately one minute, the device automatically turns
OFF.
Fig. 5
Battery Shut Down
When battery shutdown condition occurs (shown in Fig. 5):
(1)
Leave the area immediately.
(2)
Recharge or replace the battery pack.