1.2

Liability Information

MSA accepts no liability in cases where the product has been used inappropriately or not as intended.
The selection and use of this product must be under the direction of a qualified safety professional who
has carefully evaluated the specific hazards of the jobsite where it will be used and who is completely
familiar with the product and its limitations. The selection and use of this product and its incorporation
into the safety scheme of the jobsite is the exclusive responsibility of the employer.
Product liability claims, warranties also as guarantees made by MSA with respect to the product are
voided, if it is not used, serviced or maintained in accordance with the instructions in this manual.
1.3

Safety and Precautionary Measures

Carefully review the following safety limitations and precautions before placing this device in service.
Incorrect use can cause death or serious personal injury.
-
Check function (see chapter 4.8) each day before use. MSA recommends carrying out a routine
inspection prior to each day's use.
-
It is recommended that a Bump Test is performed before each day's use (see chapter 4.9) to
verify proper device operation. The device must pass the bump test. If it fails the test, perform a
calibration (see chapter 4.10) before using the device.
-
The ALTAIR 5X PID Detector is designed to detect gases and vapors in air only.
-
Bluetooth Operation is dependent upon signal availability of the wireless service(s) necessary to
maintain the communication link. Loss of wireless signal will prevent communication of alarms and
other information to linked devices. Take appropriate precautions in the event a loss of wireless
signal occurs.
It is very important to have an understanding of PID basics when changing PID settings. Failure to
properly identify the VOC gas being measured and/or failures to select the correct Response Factor
alarm values (exposure, STEL, TWA) that match the desired Response Factor and/or the correct
lamp will result in erroneous readings or erroneous alarm limits that could cause death or serious
personal injury.
Perform a Bump Test more frequently if the device is subjected to physical shock or high levels of
contaminants. Also, check calibration more frequently if the tested atmosphere contains the following
materials, which may desensitize the combustible gas sensor and/or VOC sensor (PID) and reduce its
readings:
-
Organic silicones
-
Silicates
-
Lead-containing compounds
-
Sulfur compound exposures over 200 ppm or exposures over 50 ppm for one minute
-
High concentration of VOC gas may affect CO sensor performance
-
The minimum concentration of a combustible gas in air that can ignite is defined as the Lower
Explosive Limit (LEL). A combustible gas reading of
100 % LEL, and an explosion hazard exists. Move away from hazardous area immediately.
-
Do not use the device to test for combustible or toxic gases in the following atmospheres as this
may result in erroneous readings:
-
Oxygen-deficient or oxygen-rich atmospheres
-
Reducing atmospheres
-
Furnace stacks
-
Inert environments
-
Atmospheres containing combustible airborne mists/dusts.
WARNING
WARNING
XXX
indicates the atmosphere is above
ALTAIR 5X PID
Safety Regulations
US
8