Electromagnetic immunity
The MDRC-1219 (TS) is intended for use in the electromagnetic environment (IEC 60601-1-2 4
specified below. The customer or the user of the MDRC-1219 (TS) should assure that it is used in such
an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
1. is the a.c. mains voltage prior to application of the test level.
K5902128 (451920612451) MDRC-1219 (TS) 03/11/2016
IEC 60601-1-2 4
th
edition (2014)
Test levels
± 8kV contact
± 15kV air
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
1
< 5% U
(> 95% dip in
T
U
) for 0.5 cycle
T
40% U
(60% dip in U
T
for 5 cycles
70% U
(30% dip in U
T
for 25 cycles
< 5% U
(>95% dip in
T
U
) for 5s
T
30 A/m
3 V/m @ 150 kHz to 80
MHz
9 to 28 V/m in
communication service
bands up to 6 GHz
Compliance level
± 8kV contact
± 15kV air
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5% U
(> 95% dip in
T
U
) for 0.5 cycle
T
)
40% U
(60% dip in U
T
T
for 5 cycles
)
70% U
(30% dip in U
T
T
for 25 cycles
< 5% U
(>95% dip in
T
U
) for 5s
T
30 A/m
3V/m @ 150 kHz to
80MHz
9 to 28 V/m in
communication service
bands up to 6 GHz
5. Important information
edition)
th
Electromagnetic
environment –
guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Mains power quality
should be that of a typical
commercial or hospital
environment
Mains power quality
should be that of a typical
commercial or hospital
environment
Mains power quality
should by that of a typical
commercial or hospital
)
T
environment. If the
user of the MDRC-1219
)
(TS) requires continued
T
operation during power
mains interruptions, it
is recommended that
the MDRC-1219 (TS)
be powered from an
uninterruptible power
supply or a battery.
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the MDRC-1219
(TS), including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d = 1.2√P
27