4. MAINTENANCE
4.1
Scheduled maintenance
About
The MDRC-1219 (TS) does not require any scheduled maintenance or calibration activities. We recom-
mend to use QAWeb with the Barco default tests and frequencies to calibrate and maintain the display,
or to return the display to a Barco approved maintenance organization. In any case of doubts, please
contact Barco Healthcare.
4.2
Cleaning
W
ARNING
display.
C
AUTION
rings or other jewelry and do not apply excessive pressure on the front glass or LCD.
C
AUTION
damage to internal electronics. Instead, apply the liquid to a cleaning cloth.
To clean the display
Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized clean-
ing product for medical equipment. Read and follow all label instructions on the cleaning product. In case
of doubt about a certain cleaning product, use plain water.
Possible cleaning solutions:
•
70% isopropyl alcohol
•
1.6% aqueous ammonia
•
Cidex® (2.4% glutaraldehyde solution)
•
Sodium hypochlorite (bleach) 10%
•
"Green soap" (USP)
•
0.5% Chlorehexidine in 70% isopropyl alcohol
•
Like Cleansafe® optical cleaning liquid
Do not use following products:
K5902128 (451920612451) MDRC-1219 (TS) 03/11/2016
:
Unplug the power cable from the mains power input before cleaning the
:
Take care not to damage or scratch the front glass or LCD. Be careful with
:
Do not apply or spray liquid directly to the display as excess liquid may cause
4. Maintenance
19