Measurements can be controlled via 3 freely definable measuring/saving programs. The
channel, starting criteria, measuring rate, measurement duration or number of values are
defined in the programs 1-3. The programs are started once the "starting criterion" is ful-
filled. The starting criterion is checked at the measuring rate: if a slow rate has been set it
is advisable to set a an intermediate rate when checking the starting criterion (see Chapter
8.6.5). All 3 programs can be started independently of each other, they save all of the con-
nected probe values at the set measuring rate. "Programming 1" can be taken as an exam-
ple for all 3 programmings.
Example 1: Measurement with activated "Program 1", Starting criterion: "Overrun"
Display, channel 1
Progression of parameter at channel 1
Limit value of
starting criterion
"Exceeding"
Data acquisition corresponding to the
xxxxx =
x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x
Checking the
starting criterion
Start of program 1 when
at a quicker rate
the programmed starting
("Device" menu,
criterion is entered
"Intermediate
rate" submenu)
Example 2: Measurement with two activated programs (Prog. 1 and Prog. 2),
Starting criteria: "Exceeding"
Display, channel 1
Progression of parameter at channel 1
Limit of starting
criterion
"Exceeding"
Prog. 2
Limit of
starting criterion
"Exceeding"
Prog. 1
Prog.1
active
Changing measuring rate e.g. high
resolution in critical areas
8
"Measuring rate"
Time
End of program 1, the starting
criterion is lifted
The program can be ended pre-
maturely via "Measuring rate" or
"Number of values
Time
Prog. 2
Prog.1
active
active
Main menu
8.3 Memory