Ce Approvals Information - Omega DPF700 User Manual

Hide thumbs Also See for DPF700:
Table of Contents

Advertisement

CE A
This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and
with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Double Insulation Primary to Secondaries
Pollution Degree 2
Dielectric withstand Test per 1 min
• Power to Input/Output:
• Power to Relays Output:
• RS232 to Inputs:
• Analog to Inputs:
Measurement Category I
Category I are measurements performed on circuits not directly connected to the Mains
Supply (power). Maximum Line-to-Neutral working voltage is 50V ac/dc. This unit should
not be used in Measurement Categories II, III, IV.
Transients Overvoltage Surge (1.2/50uS pulse)
• Input Power:
• Input/Output Signals:
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control and
laboratory.
• EMC Emissions Class B of EN61326
• EMC Immunity** Table 1 of EN61326
Note:
**I/O signal and control lines require shielded cables and these cables
must be located on conductive cable trays or in conduits. Furthermore,
the length of these cables should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual for
additional information.
66
I
PPROVALS
NFORMATION
2500Vac
2500Vac
No Isolation
500Vac (720Vdc)
2500V
500V