TESTO 340 Instruction Manual page 51

Flue gas analyser
Hide thumbs Also See for 340:
Table of Contents

Advertisement

Parameter
Measuring range
P abs
600...1150hPa
Temperature (NiCrNi) -40...1200°C
at 1000°C...1200°C on probe
Efficiency
0...120%
Flue gas loss
0...99,9%
Flue gas dewpoint 0...99,9°C
CO2 determination 0...CO2 max.
(Calculated
from O2)
1 Response time 90%, recommended minimum measurement duration to guarantee correct readings: 3min
For activated single dilution slot 2 (factor 5)
Parameter
Measuring range
CO, H2-comp.
700...50000ppm
COlow, H2-comp.
300...2500ppm
SO2
500...25000ppm
NO
500...15000ppm
NOlow
150...1500ppm
With activated dilution of all sensors (optional) (factor 2)
Parameter
Measuring range
O2
0...25Vol.%
CO, H2-comp.
700...20000ppm
COlow, H2-comp.
300...1000ppm
NO2
200...1000ppm
SO2
500...10000ppm
NOlow
150.. .600ppm
NO
500...6000ppm
1
Response time 90%, recommended minimum measurement duration to guarantee correct readings: 3min
Filter lifetime
Parameter
Lifetime
CO, H2-comp.
170000 ppmh
NO
120000 ppmh
Accuracy
±10hPa
± 0.5°C
at 0.0...99°C
± 0.5% of reading
-
-
-
± 0.2 Vol%
Accuracy
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
Accuracy
±1Vol.% of reading additional error (0...4,99Vol.%) 0.01Vol.%
±0,5Vol.% of reading additional error(5...25Vol.%)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
Testo-Direct
www.
Resolution
1hPa
0.1°C at -40.0...999.9°C
in rest of range
0.1%
0,1%
0.1%
0.1 Vol%
Resolution
1ppm
0.1ppm
1ppm
1ppm
0.1ppm
Resolution
1ppm
0.1ppm
0.1ppm
1ppm
0.1ppm
1ppm
1.800.561.818
.ca
51
t90
1
-
depends
0.1°C
-
-
-
<40s
1
t90
< 20s