Inolab pH level 2 Instruction Manual page 58

Laboratory ph meter
Table of Contents

Advertisement

LoBat display
to display
to
to
to
Obviously incorrect
measured values
Instrument does not
react to keystroke
You would like to
know which soft-
ware version is in
the instrument
Cause
– Batteries almost
depleted
Cause
– Timeout of the interface
Cause
pH electrode:
– pH electrode unsuitable
– Temperature difference
between buffer and
sample too large
– Measuring procedure not
suitable
Cause
– Operating state
undefined or EMC
electric stress unallowed
Cause
– e.g. question of the WTW
service department
What to do if...
Remedy
– Replace batteries
(see section
5.1 M
AINTENANCE
Remedy
– Checkout connected
instrument
Remedy
– Use suitable electrode
– Adjust temperature of
buffers or samples
– Follow special procedure
Remedy
– Processor reset:
a
Press the
key and
switch on instrument
Remedy
a
– Press the
switch on instrument.
The software version
is displayed.
)
key and
127