Data Transmission; Data Transmission Interval (Int 2) - Inolab pH level 2 Instruction Manual

Laboratory ph meter
Table of Contents

Advertisement

Operation
112
4.5

Data transmission

You can use the following options to transmit data:
l
One of the following options:
– The AutoStore function (page 106) is used to
periodically (Int 1 storage interval) save measured
values internally and output them on the interface.
– The data transmission interval (Int 2) function is used to
periodically output measured values to the interface
(see below).
l
The Output data store function (page 108) is used to out-
put calibration data and stored measured values to the
interface.
l
The analog recorder output (page 114) is used to output
measured values as voltages.
l
Data can be transmitted bidirectionally (page 116) using
the KOM pilot communication package (accessory).

4.5.1 Data transmission interval (Int 2)

The interval for the data transmission (Int 2) is determined
by the time interval between automatic data transmissions.
After the time interval expires, the current data record is
transmitted to the interface.
Note
The setting of the interval (Int 2) only has an effect when the
storage interval (AutoStore function) is switched off.