Inolab pH level 2 Instruction Manual page 30

Laboratory ph meter
Table of Contents

Advertisement

5 Thoroughly rinse the electrode with distilled water.
6 If necessary, set the temperature of the second buffer
solution by pressing
7 Submerse the electrode in the second buffer solution.
g
8 Press the
AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized, AR disap-
pears.
The sensor symbol shows the electrode evaluation
after the two-point calibration.
The value of the slope (mV/pH) appears on the dis-
play.
g
9 Press the
The value of the asymmetry (mV) appears on the dis-
play.
10 To return to the measuring mode: Press the
or continue with the three-point calibration.
u d
.
key.
key.
Operation
m
key
99