Inolab pH level 2 Instruction Manual page 23

Laboratory ph meter
Table of Contents

Advertisement

Operation
AutoRead
Calibration protocol
92
In calibration using AutoCal TEC and AutoCal DIN, the
AutoRead function is automatically activated.
The current AutoRead measurement (with acceptance of
the current value) can be terminated at any time by pressing
g
.
The calibration protocol contains the calibration data of the
current calibration. You can call up the calibration protocol
by outputting the data storage (section 4.3.2).
Note
You can automatically print out a calibration protocol after
the calibration. To do so, connect a printer according to sec-
tion 4.5.3 before the calibration. After a valid calibration, the
protocol is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.99
14:19
Device No.: 12345678
CALIBRATION pH
Cal Time: 01.03.99 / 15:20
Cal Interval:
AutoCal DIN
Tauto
Buffer 1
1.679
Buffer 2
4.008 *
Buffer 3
6.865
Buffer 4
9.180 *
C1
174.1mV
C2
-133.3mV
S1
-59.4 mV/pH
ASY1
-
4 mV
Probe:
+++
7d
25.0°C
25.0°C