Measuring
parameters
Configuration
parameters
4.7
Reset
You can reset (initialize) measuring and configuration
parameters separately from one another.
The following measuring parameters (pH InI) are reset to the
values they had on delivery:
Measuring mode
Asymmetry
Slope
Calibration procedure
Temperature, manual
Resolution of pH display
Note
When the measuring parameters are reset, the calibration
data are lost. After the parameters have been reset, cali-
brate!
The following configuration parameters (InI) are reset to the
values they had on delivery:
Baud rate
Interval 1
(automatic storing)
Interval 2
(for data transmission)
pH
0 mV
-59.16 mV/pH
AutoCal TEC
25°C
0.01
4800
OFF
OFF
Operation
121