Single-point
calibration
Use a buffer solution in the range pH = 7.0 ± 0.5 for this pro-
cedure.
Note
Only the electrode asymmetry is determined in single-point
calibration. The slope of the last two-point calibration is
retained.
Note
Step 2 is not required if you use a temperature probe. The
TP message indicates an active temperature measurement.
c
1 Press the
tion display appears.
2 Set the temperature of the buffer solution by pressing
u d
.
3 Submerse the pH electrode in the buffer solution.
g
4 Press the
The measured pH value appears on the display.
5 Set the nominal pH value of the buffer solution (at the
current temperature) by pressing the
g
6 Press the
The value of the asymmetry (mV) and the sensor
symbol for the evaluation of the electrode appears on
the display.
7 To return to the measuring mode: Press the
key repeatedly until the ConCal func-
key.
key.
Operation
u d
keys.
m
key.
103