Inolab pH level 2 Instruction Manual page 18

Laboratory ph meter
Table of Contents

Advertisement

Temperature probe
Measurements can be performed with and without a temper-
ature probe. A connected temperature probe is indicated by
TP on the display.
Note
The pH meter automatically recognizes the type of the tem-
perature probe used. As a result, you can connect elec-
trodes with the NTC30 or Pt1000.
The temperature measurement is absolutely essential for a
reproducible pH measurement. If the measurement is made
without a temperature probe, proceed as follows:
1 Determine the current temperature using a thermo-
meter.
2 Set up the temperature by pressing
Note
When calibrating without a temperature probe, set up the
current temperature of the respective buffer solution manu-
ally by pressing the
u d
keys.