Table 5-10
Common Mode Stimulus Preset TDR/TDT Configuration
Menu
Acquisition
Time Base
Channel 1 Setup
Channel 2 Setup
Key Path
TDR/TDT Setup Preset TDR/TDT
Menu Item
Set To
Averaging
on
Best
Flatness
Scale
500.0 ps/div
Position
Set to a value which places the
incident edge on screen.
Display
on
Scale
100 mV/div
Offset
200.0 mV
Bandwidth
12.4 GHz (for Stimulus on)
18.0 GHz (for Stimulus external)
Attenuation units
ratio
Attenuation
1.000 : 1
Scale
100 mV/div
Offset
200.0 mV
HP 54754A TDR/TDT Setup Menu