9
Figure 7-24
1
2
3
4
5
6
7
8
9
Single-ended TDR Measurements
Measuring the Distance to a Discontinuity
Change the
until the display is similar to Figure 7-20.
Position
Press
on the TDR plug-in module.
TDR/TDT Setup
Press the
Normalize response . . .
Press the
Establish normalization & ref plane
Connect an SMA short to the end of the cable.
Remove the short from the end of the cable and connect an SMA 50 ohm
load to the end of the cable.
Press the
Continue
softkey.
Press the
Done
softkey.
Remove the 50 ohm load from the end of the cable.
Connect the channel 1 cable to the single transmission line connector
closest to the narrow trace on the demo board (Figure 7-25).