HP 54753A User Manual page 137

Plug-in modules
Table of Contents

Advertisement

8
Press the
9
Press the
10
Change the
wide trace discontinuity.
11
Read the Y voltage value at the bottom of the display.
Figure 7-18
Wide Trace
Voltage
The measured voltage (v) is 155.20 mV. Substituting into the equation:
We will now have the oscilloscope make this measurement for us. This requires
establishing normalization and the reference plane as follows.
1
Press the
2
Press the
Press the
3
4
Press the
5
Press the
Remove the cable end connected to the demo board.
6
7
Press the
Measuring Transmission Line Percent Reflection
softkey to select
Vertical units
Done
softkey.
+ Position
until the + marker is over the negative peak of the
155.20 200
×
p
=
100
------------------------------ -
max
200
+ Source
softkey and select
softkey.
Enter
key.
SETUP Math
Display
softkey to turn the function 2 display off.
key.
TDR/TDT Setup
Normalize response . . .
Single-ended TDR Measurements
.
Volt
=
22.40 %
.
response 1
softkey.
7-29

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

54754a tdr

Table of Contents