HP 54753A User Manual page 166

Plug-in modules
Table of Contents

Advertisement

4
Figure 8-12
Measured Skew
1
2
3
4
Differential TDR Measurements
Making Differential TDT Measurements
Change the
until the reflected edge is approximately centered
Position
in the display (Figure 8-12).
A ∆time (skew) value of 7 ps (one fifth the TDR step risetime of 35 ps) or less
is small enough that the skew will not introduce errors into TDR measurements.
However, for the purposes of demonstration, the following process shows how
to deskew the cables.
Press the
SETUP Channe
right-most step on the display.
Press the
Calibrate . . .
Change the
until the
Skew
Change the
TDR Skew
will need to move the control slowly as it takes a short amount of time
for the waveform to settle (Figure 8-13).
8-18
l key for the channel whose reflected step is the
softkey.
Time is ½ its initial value.
until the remaining
Time is approximately 0. You