Establish Ref Plane - HP 54753A User Manual

Plug-in modules
Table of Contents

Advertisement

Response scaling . . .
Key Path
Vertical There are two choices for vertical mode: track source and manual .
The track source mode sets the control of the vertical scaling for the
normalized trace to that of the source channel. When manual mode is
selected, the Scale and Offset menus appear allowing you to independently
change the vertical scale and offset of the normalized trace.
Key Path
Horizontal There are two choices for horizontal mode: track source and
manual . The track source mode sets the control of the horizontal scaling
for the normalized trace to that of the source channel. When manual mode
is selected, the Scale and Position menus appear allowing you to
independently change the horizontal scale and position for the normalized
trace.
Key Path
HP 54754A TDR/TDT Setup Menu

Establish ref plane

The Response scaling function is used when vertical and horizontal scaling of
the normalized response is required that is independent from that of the source
channel. This function only appears when TDT response 2 is set to differential
or common mode.
TDR/TDT Setup TDT response 2 Response scaling . . .
TDR/TDT Setup TDT response 2 . . .Response scaling . . . Vertical
TDR/TDT Setup TDT response 2 . . .Response scaling . . . Horizontal
Establish ref plane
The Establish ref plane function is used to set the reference plane for TDR and
TDT measurements for differential or common mode stimulus. The function
steps you through the calibration procedure for the selected measurement type.
The steps are displayed at the top of the screen. The items required for
calibration are shown in Table 5-8.