Precision: The precision of CareSens N Test strip was estimated
with venous blood sample in the laboratory.
Within Run Precision
Blood average
38.1 mg/dL (2.1 mmol/L)
Blood average
86.1 mg/dL (4.8 mmol/L)
Blood average
124.5 mg/dL (6.9 mmol/L)
Blood average
189.1 mg/dL (10.5 mmol/L) CV = 2.6%
Blood average
334.5 mg/dL (18.6 mmol/L) CV = 2.8%
Total Precision
Control average 43.1 mg/dL (2.4 mmol/L)
Control average 113.4 mg/dL (6.3 mmol/L)
Control average 381.2 mg/dL (21.2 mmol/L) CV = 4.2%
This study shows that there could be variation of up to 4.2%

44 www.i-sens.com

SD = 1.9 mg/dL (0.11 mmol/L)
SD = 3.2 mg/dL (0.2 mmol/L)
CV = 4.1%
SD = 2.0 mg/dL (0.11 mmol/L)
CV = 3.3%
Warranty Information
Manufacturer's Warranty
i-SENS, Inc. warrants that the CareSens N Meter shall be free of
defects in material and workmanship in normal use for a period
of five (5) years. The meter must have been subjected to normal
use. The warranty does not cover improper handling, tampering,
use, or service of the meter. Any claim must be made within the
warranty period.
The i-SENS company will, at its discretion, repair or replace a
defective meter or meter part that is covered by this warranty.
As a matter of warranty policy, i-SENS will not reimburse the
consumer's purchase price.
Obtaining Warranty Service
To obtain warranty service, you must return the defective meter
or meter part along with proof of purchase to your nearest
i-SENS Authorized Warranty Station.