Setting the 'Test Result Reset'
(Deleting all the saved test results)
Step 10
In this mode all the test results stored in the meter can be
deleted. Please note that if you select YES, all the stored test
results will be deleted and can not be restored.
After the beeper mode is set,
press the S button to enter
the 'Test Result Reset' mode.
The 'dEL' character will blink
on the screen. Press or
to alternate between 'YES' or
'no' . To delete all the stored
test results press the S button
while the screen displays
'YES' . Then, all the test results stored in the meter will be deleted
and the screen will be similar to the picture on the right.
If you do not want to delete the results press the S button while
the screen displays 'no' . Then, the screen will return to step 2.
See page 12.
Note:
At any stage, if the S button is pressed for 3 seconds, Time, Date and
Unit setting mode will finish and the meter will be turned off. Press
and hold to scroll through numbers quickly.
16 www.i-sens.com
Checking the System
You may check your meter and test strips using the
CareSens Control Solution.
The CareSens Control Solution contains a known
amount of glucose and is used to check that the meter
and the test strips are working properly.
The test strip vials have CareSens Control Solution
ranges printed on their labels. Compare the result
displayed on the meter to the CareSens Control
Solution range printed on the test strip vial.
Before using a new meter or a new vial of test strips, you may
conduct a control solution test following the procedure on page 18.
Note:
• Use only the CareSens Control Solution.
• Check the expiration dates printed on the vial. When you first
open a control solution vial, record the discard date (date
opened plus 3 months) in the space provided on the label.
• Make sure your meter, test strips, and control solution are at
room temperature before testing. Control Solution tests must
be done at room temperature (20 ~ 25 °C, 68 ~ 77 °F).
• Before using the control solution, shake the vial, discard the
first few drops and wipe the tip clean.
• Close the control solution vial tightly and store at a
temperature of 8 ~ 30 °C (46 ~ 86 °F).
You may do a control solution test:
• When using the meter for the first time
• Whenever you open a new vial of test strips
• If the meter or test strips do not function properly
• If your symptoms are inconsistent with the blood glucose test
results and you feel that the meter or test strips are not working
properly
• If you drop or damage the meter