Beckman Coulter UniCel DxC Synchron 800 Instructions For Use Manual page 422

Synchron clinical systems
Table of Contents

Advertisement

Calibration Errors
Calibration Error Flags
Table 13.1
Printed Flag
OCR LOW
RECOVERY
SENSITIVITY
SEV RECOVERY
SEV SENSITVITY
SPAN
RANGE
Troubleshooting Calibration and Result Errors
Page 13-2
Calibration Error and Remarks
Remarks on Report
Calibration failed
Calibration failed
Calibration failed
Calibration required
Calibration required
Calibration failed
Calibration failed
, continued
Description
Response out of range
low.
Recovery error
Sensitivity error
Severe recovery error
Severe sensitivity error
Sensitivity error
Accuracy error (ISE)
UniCel DxC Systems Instructions For Use A13914