Monthly Maintenance
Replace CTS Blade (Sarstedt S-Monovette Type CTS Option only)
Step
3
• Fully loosen the two captive Cap Piercer cover screws (1) at the base of
• Lift and remove the cover.
Maintenance
Page 10-34
The points of the Blade are very sharp and extend below the Wash
Tower. Stay away from the bottom of the Wash Tower.
DO NOT select <OK> when the cover is removed. This would move the
Cap Piercer Carriage up and cause a pinching hazard.
the Cap Piercer cover.
Figure 10.18
Action
, continued
CAUTION
CAUTION
1
UniCel DxC Systems Instructions For Use A13914
A014462P.EPS
(2 of 4)
October 2005