Enzyme Validator

Introduction
The Enzyme Validator option provides the ability to obtain IFCC/DGKCh (refer to
"Additional
Information" below) equivalent answers for selective enzymes, or to
obtain IFCC/DGKCh equivalent answers for the non-IFCC/non-DGKCh formulation
enzymes.
• The Synchron Enzyme Validator set is a human serum, albumin-based value
assigned material and is used like a calibrator.
• The feature can be turned on or off for applicable chemistries; the default is off.
• The Enzyme Validator Kit includes diskettes which provide set points for the
selective enzymes.
Once the enzyme reagent is configured and loaded, and the Enzyme Validator
feature is turned on, Enzyme Validator ID assignments and calibration requests
can be performed in the same manner as any calibrator.
Step
1
Select the Rgts/Cal icon from the menu bar.
2
Select Options [F6].
3
From the Cal Options dialog box, select Enzyme Validator.
4
Using the pull-down menu, select a desired Configured Chemistry name.
5
Select <OK> to activate the Enzyme Validator option.
Select <Cancel> to exit without enabling the option.
Additional Information
DGKCh (Deutsche Gesellschaft für Klinische Chemie - German Clinical Chemistry
Association), a committee that establishes formulation standards in Europe.
IFCC (International Federation of Clinical Chemistry)
UniCel DxC Systems Instructions For Use A13914
October 2005