USP MODE PROCEDURES

USP MODE PROCEDURES
USP MODE PROCEDURES
USP MODE PROCEDURES
USP MODE PROCEDURES
Press Mode key while in EC range to enter USP mode.
The instrument will display the USP main screen.
Press Stage 1 to start with first stage evaluation.
Press Stage 2 to start with second stage evaluation.
If Stage 1 is pressed a tutorial screen is displayed.
Use the Arrow keys to scroll the tutorial message.
Press Continue to skip the tutorial message and enter EC measuring mode.
If the temperature source is manual entry, press USP Fact. to select USP factor
to be changed (the is displayed before the 100% value). To increase the
accuracy of the analyze decrease the USP factor value.
Wait until reading is stable.
If the sample reading met the USP stage 1 criteria, the "USP Met" message
will be displayed.
Press Report to view report.