According to the definition, salinity of a sample in psu (practical salinity units)
is calculated using the following formula:
R
- coefficient;
T
C
(sample) - uncompensated conductivity at T °C;
T
C(35,15)= 42.914 mS/cm - the corresponding conductivity of KCl solution
containing a mass of 32.4356 g KCl / 1 Kg solution;
r
- temperature compensation polynom
T
a
= 0.008
b
=0.0005
0
0
a
= -0.1692
b
= -0.0056
1
1
a
= 25.3851
b
= -0.0066
2
2
a
= 14.0941
b
= -0.0375
3
3
a
= -7.0261
b
= 0.0636
4
4
a
= 2.7081
b
= -0.0144
5
5
c
= 0.008
0
c
= 0.0005
1
X = 400R
T
Y = 100R
T
f (T)= (T-15)/[1+0.0162(T-15)]
Note:
• The formula can be applied for salinity values between
0 and 42 psu.
• The formula can be applied for temperatures between
-2 °C and 35 °C.
To reach this range press Mode while in Salinity range until the practical
salinity scale [PSU] is displayed.