A U T O L O G
A U T O L O G
A U T O L O G
A U T O L O G
A U T O L O G
This feature allows the user to log up to 1000 readings.
All logged data can be transferred to PC through USP port.
The memory space is organized in lots of records. A lot can contain from 1 to
1000 records. The maximum available lots number is 100.
START AUTOLOG
Set the desired logging interval in the Setup menu.
If 5, 10, 30 s or 1 min interval is selected, the AutoPower Off feature is
disabled (each sample memorized restarts the autopower off counter). For
other intervals, the instrument will enter a sleeping mode. During this mode
the instrument continues to monitor the inputs and memorize the reading at
the set interval. To exit sleeping mode, simply press any key (exception ON/
OFF).
Note: While in sleeping mode the instrument cannot be powered off by
pressing ON/OFF key. Leave sleeping mode and then press ON/OFF
key to power off the instrument.
To start autolog from the measurement screen press StartLog key. The lot
number and the amount of memory space is displayed for few seconds.
If the autolog space is full or the lots number exceeds 100 the
"Auto log space full" message is displayed.
VIEW AUTOLOG DATA
Press RCL key to enter View Logged Data mode.