Hanna Instruments HI98194 Instructions Manual

Hanna Instruments HI98194 Instructions Manual

Multiparameter meters ph/mv, orp, ec, tds, resistivity, salinity, seawater, dissolved oxygen, atmospheric pressure and temperature
Hide thumbs Also See for HI98194:

Advertisement

HI98194,
HI98195, HI98196
Multiparameter Meters
pH/mV, ORP , EC, TDS, Resistivity,
Salinity, Seawater σ, Dissolved
Oxygen, Atmospheric Pressure
and Temperature
sale@ponpe.com
www.ponpe.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI98194 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI98194

  • Page 1 HI98194, HI98195, HI98196 Multiparameter Meters pH/mV, ORP , EC, TDS, Resistivity, Salinity, Seawater σ, Dissolved Oxygen, Atmospheric Pressure and Temperature sale@ponpe.com www.ponpe.com...
  • Page 2 All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA. sale@ponpe.com www.ponpe.com...
  • Page 3: Table Of Contents

    CHAPTER 1 - INTRODUCTION Preliminary Examination ..................... Model Identification ......................General Description ......................Display and Keypad Description ................... CHAPTER 2 - QUICK START Sensor and Probe Installation ....................Basic Operation ......................... Help Function ........................CHAPTER 3 - SPECIFICATIONS System Specifications ......................Probe Specifications ......................
  • Page 4 CHAPTER 7 - CALIBRATION MODE Calibration Mode......................... Quick Calibration ........................ pH Calibration ........................Relative mV Calibration ....................... Dissolved Oxygen Calibration (HI98194, HI98196 only) ............Conductivity Calibration (HI98194, HI98195 only) ..............Temperature Calibration ...................... Atmospheric Pressure Calibration ..................CHAPTER 8 - SYSTEM SETUP Meter Setup ........................
  • Page 5: Chapter 1 - Introduction

    Note: Save all packing materials until you are sure that the instrument functions correctly. Any damaged or defective items must be returned in their original packing material with the supplied accessories. pH/mV, ORP, EC, TDS, Resistivity, Salinity, Seawater σ, Dissolved Oxygen, HI98194 Atmospheric Pressure and Temperature Multiparameter meter with HI7698194 probe.
  • Page 6: General Description

    • Measure up to 14 parameters and display of up to 12 parameters • Waterproof protection (IP67 for the meter and IP68 for the probe) • Graphic LCD with backlight • Built‑in barometer for D.O. concentration compensation (HI98194, HI98196 only) •...
  • Page 7: Display And Keypad Description

    1. Graphic LCD 2. Battery level indicator 3. Softkeys On/Off key: turn the meter on and off Lamp key: turn the backlight on and off 6. Alphanumeric keyboard: insert alphanumeric codes 7. / Arrow keys: scroll the displayed options/message 8. HELP key: obtain information about the displayed screen 9.
  • Page 8: Chapter 2 - Quick Start

    HI769819X probes have 2 or 3 sensor connectors sockets identified with color‑coded triangles: • Connector 1 (red): For either pH/ORP, pH sensor • Connector 2 (white): For dissolved oxygen sensor (HI98194, HI98196 only) • Connector 3 (blue): For EC sensor (HI98194,...
  • Page 9 sale@ponpe.com www.ponpe.com...
  • Page 10: Basic Operation

    The main operating modes for HI9819X are measurement, logging and setup. The measurement screen can be configured to display a single measurement or up to 12 simultane‑ ous measurements by using the numbers 1‑7 on the keypad. Use the / keys to scroll through the measurements not being displayed.
  • Page 11: Chapter 3 - Specifications

    0.1 mV Accuracy ± 1.0 mV Calibration Automatic at 1 custom point (relative mV) DISSOLVED OXYGEN (HI98194, HI98196 only) Range 0.0 to 500.0 %; 0.00 to 50.00 ppm (mg/L) Resolution 0.1 %; 0.01 ppm (mg/L) 0.0 to 300.0 %: ± 1.5 % of reading or ± 1.0 % whichever is greater;...
  • Page 12 Depending on resistivity reading Calibration Based on conductivity or salinity calibration TDS (Total Dissolved Solids) (HI98194, HI98195 only) Range 0 to 400000 ppm (mg/L); (the maximum value depends on the TDS factor) Manual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L);...
  • Page 13 ±2% of reading or ±0.01 PSU whichever is greater Calibration Based on conductivity calibration Note: For HI98196 Salinity can be set from 0.00 to 70.00 PSU in setup menu. SEAWATER SIGMA (HI98194, HI98195 only) 0.0 to 50.0 σ σ σ...
  • Page 14 METER SPECIFICATIONS Temperature Compensation Automatic from ‑5 to 55 °C (23 to 131 °F) 45,000 records (continuous logging or log‑on‑demand of all Logging Memory parameters) Logging Interval 1 second to 3 hours PC Interface USB (with HI9298194 software) Waterproof Protection IP67 Environment 0 to 50 °C (32 to 122 °F);...
  • Page 15: Probe Specifications

    3 for HI7698194 Sensor Inputs 2 for HI7698195 HI7698196 Sample Environment Fresh, brackish, seawater Waterproof protection IP68 Operating Temperature ‑5 to 55 °C Storage Temperature ‑20 to 70 °C Maximum Depth 20 m (66’) Dimensions 342 mm (13.5”), dia=46 mm (1.8”) (without cable) Weight (with 570 g (20.1 oz.)
  • Page 16: Sensor Specifications

    HI7698194‑0 HI7698194‑1 HI7698194‑2 HI7698194‑3 Description pH/ORP Dissolved Oxygen Measure Type D.O. pH, mV (pH) pH, mV (pH/ORP) Primary Unit (% sat. & conc.) 0.00 to 13.00 pH 0.0 to 200.0 mS/cm Measure 0.00 to 13.00 pH 0.0 to 500.0 % ±600.0 mV 0.0 to 400 mS/cm Range...
  • Page 17: Chapter 4 - Probe Installation

    Chapter 4 - PROBE INSTALLATION HI7698194‑0 Combination pH sensor features a glass pH sensitive bulb and a silver/silver chloride double junction reference with gelled electrolyte. HI7698194‑1 Combination pH/ORP sensor features a glass sensitive bulb for pH readings, a platinum sensor for redox measurements and a silver/silver chloride double junction reference with gelled electrolyte.
  • Page 18: Sensor Preparation/Activation

    pH Preparation Remove the shipping cap from the pH sensor. If the shipping cap does not contain any liquid, pour HI70300 into shipping cap, place it back on the sensor and soak for at least 1/2 hour before use. If HI70300 is not available, pH 4.01 buffer may be substituted.
  • Page 19: Sensor Installation

    EC Sensor Preparation The EC sensor does not need to be soaked or hydrated before use. Use the small brush included in the probe maintenance kit to clean and loosen any debris before using. HI7698194 can support three different sensors: Connector 1: pH, pH/ORP Connector 2: D.O.
  • Page 20 HI7698195 support two different sensors: Connector 1: pH, pH/ORP Connector 2: EC HI7698196 support two different sensors: Connector 1: pH, pH/ORP Connector 2: D.O. sale@ponpe.com www.ponpe.com...
  • Page 21 For a correct installation: • Grease the sensor O‑Ring with the lubricant found in the probe maintenance kit. DO NOT SUBSTITUTE other grease/lubricants as it may cause the O‑Ring to swell. • Insert the sensor into the correctly color coded opening while positioning the connector key toward the center of the probe.
  • Page 22: Chapter 5 - Initialization And Measurement

    Chapter 5 - INITIALIZATION AND MEASUREMENT HI9819X are supplied with 4 alkaline, size AA batteries. The battery symbol on the LCD indicates the remaining battery life. The meter has a low battery warning, and when the symbol starts blinking, batteries should be replaced with new ones. When the batteries are discharged the meter will automatically shut off to avoid erroneous readings.
  • Page 23: Meter Initialization

    The compatible meter‑probe pairs are: HI98194 ‑ HI7698194; HI98195 ‑ HI7698195; HI98196 ‑ HI7698196. If an incompatible sensor was connected to the probe, “Incompat.sensor!” will be displayed according to the respective connector.
  • Page 24: Measurement Mode

    Two active soft keys are found at the bottom of the status screen. • Press Measure to access the measurement mode. • Press Param to access the “Select Parameter” menu. (This screen can also be accessed from the main menu, see Chapter 6 for a detailed description.). •...
  • Page 25: Setup Menu Structure

    sale@ponpe.com www.ponpe.com...
  • Page 26: Chapter 6 - Parameter Setup Menu

    - ppt. D.O. Concentration Unit (HI98194, HI98196 only) The user can select ppm or mg/L. Dissolved Oxygen concentration is calculated using % saturation, conductivity (HI98194) or salinity constant (HI98196) and atmospheric pressure. The default value is ppm.
  • Page 27 1μS/cm, 0.001 mS/cm, 0.01mS/cm, 0.1mS/cm or 1mS/cm: the meter will not autorange, the measurement will be displayed with the selected resolution. The default value is Auto. Absolute EC Resolution Unit (HI98194, HI98195 only) Absolute conductivity displays the conductivity without temperature compensation. See EC resolution for resolution details.
  • Page 28: Parameter Coefficients

    EC Reference Temperature (HI98194, HI98195 only) This value is used for temperature compensated conductivity. All EC measurements will be referenced to the conductivity of a sample at this temperature. Press the softkey to select the desired option; 20 °C or at 25 °C. The default value is 25 °C.
  • Page 29: Calibration Mode

    This value should be kept low if you want a fast response. Press Modify to select the desired number of samples to average. This value can be set from 1 to 20 samples. The default value is 1. Note: Each reading takes 1 second, so when logging the first sample will be delayed by a few seconds if averaging is used.
  • Page 30: Quick Calibration

    The quick calibration method provides a quick single point calibration for pH, conductivity and dissolved oxygen sensors. HI9828‑0 calibration solution is used for both pH and conductivity. • Fill the calibration beaker 2/3 full with HI9828‑0 calibration solution. • Slowly place the sensors into the solution and dislodge bubbles that may adhere to the sensors. •...
  • Page 31 Note: Do not attempt to dry wipe the D.O. sensor as damage to the membrane may occur. • Screw the empty calibration beaker on the probe body. The beaker should not be dry. • Press Accept to close the displayed message. •...
  • Page 32 Procedure The measured pH value is displayed, along with the temperature and the buffer value on the second level. If necessary, press the Cal point softkey and use the / keys to select the correct buffer. • Immerse the sensors in the first buffer rinse solution and stir gently. •...
  • Page 33: Ph Calibration

    pH Calibration Error Messages HI9819X displays a series of messages if an error has occurred during calibration. If the meter does not accept a pH calibration point, a short message is displayed to indicate the possible error source. The following screens are examples: These are the available messages: •...
  • Page 34: Relative Mv Calibration

    The “ORP calibration” allows the user to perform a single point custom calibration (relative mV) or to restore the factory calibration. The Oxidation‑Reduction Potential (ORP), displayed in mV, is the voltage that results from the difference in potential between the platinum ORP sensor and the silver/silver chloride reference electrode.
  • Page 35: Dissolved Oxygen Calibration (Hi98194, Hi98196 Only)

    • To restore the factory calibration data, select the corresponding option in the “ORP calibration” menu and then press Select. DISSOLVED OXYGEN CALIBRATION (HI98194, HI98196 only) The accuracy of dissolved oxygen measurements is directly related to membrane cleanliness and calibration technique. Oily coating and biological contaminants are the primary cause of calibration drift in dissolved oxygen sensors.
  • Page 36 The calibration of the D.O. concentration range can only be performed at a single custom point (4 to 50 mg/L). It is recommended to calibrate the D.O. sensor close to the values that will be measured. Choose “DO calibration” from the “Calibration” menu, select the D.O. calibration type using the / keys and press Select to confirm.
  • Page 37 • Press ESC twice to return to the main menu. • Press Measure to return to the measurement screen. D.O. concentration (HI98194, HI98196 only) Verify the barometric pressure, conductivity and temperature reading are correct. Calibrate them if necessary. To calibrate the D.O.
  • Page 38: Conductivity Calibration (Hi98194, Hi98195 Only)

    A conductivity calibration is used to adjust for variations in cell factors by using a standard solution of known conductivity. Oily coating and biological contaminants are the primary cause of calibration drift in conductivity sensors. This type of fouling changes the apparent cell geometry, resulting in a shift in cell constant. Before performing a conductivity calibration inspect the EC sensor for debris or blockages.
  • Page 39 • Immerse the sensor into the rinse standard by raising and lowering the beaker a few times to ensure that the EC sensor channels are filled with fresh standard. • Place the calibration beaker over the EC sensor and dislodge any trapped bubbles. Screw the beaker into place.
  • Page 40 • Place the calibration beaker over the EC sensor and dislodge any trapped bubbles. Screw the beaker into place. Wait for the reading to stabilize. The stability timer will count down and Confirm will appear. • Note the temperature and adjust the conductivity value if needed. •...
  • Page 41: Temperature Calibration

    Notes: These procedures calibrate the slope value. To calibrate the offset, set the calibration point at 0 μS/cm and repeat the procedure. If the temperature input is not within the acceptable range (0 to 50 °C), the message “Invalid temperature” is displayed.
  • Page 42: Atmospheric Pressure Calibration

    Place HI9819X in a wind‑free area and choose “Custom pressure” to perform a user calibration or “Restore factory calib”. Note: “Custom pressure” procedure requires a reference barometer. Select the “Atm. pressure” from the “Calibration” menu. • Select the “Custom pressure” option. •...
  • Page 43 Date The date and date format are set in this function. Press Modify and set the date using the keypad. Press Accept to save the date. Press Format to change between the available date formats: DD/MM/YYYY, MM/DD/YYYY, YYYY/MM/DD, YYYY‑MM‑DD, MM‑DD‑YYYY, and DD‑MM‑YYYY. The default format is YYYY/MM/DD. Auto Poweroff The Auto Poweroff function is used to save battery life.
  • Page 44 Error Beep If enabled, an acoustic signal sounds every time an incorrect key is pressed, or when an error occurs. A checked box indicates this function has been enabled. The default setting is disabled. Decimal Separator The user can select the type of decimal separator: “dot” or “comma”.
  • Page 45: Probe Setup

    • The meter will require password confirmation. Retype the same password and press Accept to confirm. • The meter returns to the “Meter Setup” menu. The checkbox corresponding to the meter password is checked. To disable the password protection highlight “Meter Password” and press Modify, enter the password and then press Disable.
  • Page 46: Chapter 9 - Status

    Chapter 9 - STATUS Useful information regarding the meter, probe (if connected) and GLP calibration data are available for viewing by selecting “Status” from the main menu. Select “Meter Status” to display information related to the battery, logging, internal temperature, password, Meter ID, serial number and firmware version.
  • Page 47: Glp Data

    GLP (Good Laboratory Practice) is a set of functions that allows the user to store or recall data regarding the probe calibration. This feature also allows the user to associate readings with specific calibrations. To view GLP data select “GLP” from the “Status” menu. The complete list of available parameters appears.
  • Page 48 • From the “GLP” menu select the “ORP” option. • Data regarding the last ORP calibration will be displayed: calibration point, time and date. • Use the / keys to scroll through the stored data for the last 5 calibrations. •...
  • Page 49: Logging Mode

    Temperature • From the “GLP” menu select the “Temperature”option. • Data regarding the last temperature calibration will be displayed: calibrated point, time and date. • Use the keys to scroll through the stored data for the last 5 calibrations. Notes: If no user temperature calibration has been performed or if calibration was cleared using the “Restore factory calib.”...
  • Page 50: Logging Menu Structure

    From measurement mode, press Log to access the log menu. • The data logged on the meter are organized by lots. Up to 45,000 complete records can be stored in up to 100 lots. Each lot can store log‑on‑demand records and/or continuous records with different parameter configurations.
  • Page 51 One Sample On Meter Use this option to log one set of enabled measurement parameters to the meter memory. • If there are no lots saved on the meter, press New to create a new lot. Use the keypad to enter the desired lot name and press Accept to confirm.
  • Page 52: Log Recall

    • Select “Meter log recall” to view logs that are stored on the meter. The meter will show the number of available lots. Select “Lots” to view or delete individual lots. • Use the / keys to select the desired lot and then press View.
  • Page 53: Log Notes

    • Use the / keys to move the cursor in the graph and highlight a sample. The sample data are displayed below the graph. • Press ESC to return to the parameter list. • Press ESC again to return to the menu. Note: The number of lot samples that can be plotted is limited by the display resolution.
  • Page 54: Pc Connection

    Chapter 11 - PC CONNECTION MODE The logged data from a probe or meter can be transferred to a PC using the HI9298194 Windows compatible application software. HI9298194 offers a variety of features and on‑line‑help is ® available. HI9298194 allows data to be imported into most spreadsheet programs (e.g. Excel ©...
  • Page 55 Meter to PC data a. Select parameter units b. Select Meter from toolbar c. Select Lot sale@ponpe.com www.ponpe.com...
  • Page 56: Chapter 12 - Troubleshooting / Error Messages

    Chapter 12 - TROUBLESHOOTING / ERROR MESSAGES HI9819X displays error messages to aid in troubleshooting. Warnings are displayed for most issues, while Errors are displayed for critical issues. See the calibration chapter for messages that can occur during calibration. Other messages are listed below. •...
  • Page 57: Appendixa - Probe Maintenance

    HI7698292 probe maintenance kit includes HI7042S (electrolyte solution for D.O. sensor), spare membranes with O‑Rings for D.O. sensor, a small brush for cleaning EC , O‑Rings for sensor connectors and a syringe with grease to lubricate these O‑Rings. General Maintenance •...
  • Page 58 After cleaning soak the sensor in HI70300 storage solution before taking measurements. D.O. Sensor Maintenance (HI98194, HI98196 only) For a top performance probe, it is recommended to replace the membrane every 2 months and the electrolyte monthly. Proceed as follows: •...
  • Page 59: B - Probe Deployment

    The Hanna HI769819X have been designed for a variety of water quality measurements both in situ or in active deployments in urban or natural waters. HI9819X systems may be used for discrete spot sampling with log on demand function or, continuous monitoring and logging.
  • Page 60 The probe is suitable for measurements in a flow cell. Pumping water to a flow‑through monitoring station has obvious pros and cons. Typically a shelter is required to secure a pump, and flow chamber. A power requirement, shelter, pump maintenance and higher installation cost need to be considered. Freeze protection, security, and convenience of calibration and possibility of adding multiple mea‑...
  • Page 61: C - Accessories

    PROBES packed in carton box, without sensors or protective shield. Code Description HI7698194/4 HI7698194 probe with 4 m (13.1’) cable. HI7698194/10 HI7698194 probe with 10 m (33.0’) cable HI7698194/20 HI7698194 probe with 20 m (65.6’) cable HI7698194/40 HI7698194 probe with 40 m (131.2’) cable HI7698195/4 HI7698195 probe with 4 m (13.1’) cable...
  • Page 62 METERS WITH PROBES Code Description HI7698194 HI98194 meter, probe with 4 m (13.1’) cable, with pH/ORP, EC, D.O. sensors. HI7698194/10 HI98194 meter, probe with 10 m (33.0’) cable, with pH/ORP, EC, D.O. sensors. HI7698194/20 HI98194 meter, probe with 20 m (65.6’) cable, with pH/ORP, EC, D.O.
  • Page 63 SENSORS Code Description HI7698194-0 pH sensor HI7698194-1 pH/ORP sensor HI7698194-2 Dissolved Oxygen sensor HI7698194-3 EC sensor CABLES, CONNECTORS, ACCESSORIES Code Description HI7698290 Short calibration beaker HI7698295 Short protective shield HI7698297 Long, quick release flow cell HI7698292 Probe maintenance kit with HI7042S (electrolyte solution for D.O.
  • Page 64 QUICK CALIBRATION SOLUTIONS Code Description HI9828-20 Quick calibration solution, 230 mL HI9828-25 Quick calibration solution, 500 mL HI9828-27 Quick calibration solution, 1 gal. pH BUFFERS HI5004 pH 4.01 buffer solution, 500 mL HI5005 pH 5.00 buffer solution, 500 mL HI5006 pH 6.00 buffer solution, 500 mL HI5068 pH 6.86 buffer solution, 500 mL...
  • Page 65 ORP SOLUTIONS Code Description HI7020L ORP test solution, 200/275 mV @ 25 °C, 500 mL HI7021L ORP test solution, 240 mV @ 25 °C, 500 mL HI7022L ORP test solution, 470 mV @ 25 °C, 500 mL HI7091L Reducing pretreatment solution HI7092L Oxidizing pretreatment solution, 500 mL pH/ORP MAINTENANCE SOLUTIONS...
  • Page 66 CONDUCTIVITY STANDARD SOLUTIONS Code Description HI7030L 12880 µS/cm calibration solution, 500 mL HI7031L 1413 µS/cm calibration solution, 500 mL HI7033L 84 µS/cm calibration solution, 500 mL HI7034L 80000 µS/cm calibration solution, 500 mL HI7035L 111800 µS/cm calibration solution, 500 mL 5000 µS/cm calibration solution, 500 mL HI7039L sale@ponpe.com...
  • Page 67 When shipping any instrument, make sure it is properly packed for complete protection. Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice. sale@ponpe.com...
  • Page 68 World Headquarters Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com Local Office Hanna Instruments USA 270 George Washington Highway Smithfield, RI 02917 Phone: 800.426.6287 Fax: 401.765.7575 e‑mail: tech@hannainst.com MAN98194 03/16 Printed in ROMANIA sale@ponpe.com...