Remove the meter from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is evident, contact your local Hanna Instruments Office. Each meter is supplied with:
Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packaging with the supplied accessories.
The pH electrode is shipped dry. Before using the pH electrode, remove the protective cap and condition the electrode by soaking the tip (bottom 3 cm (1.18")) in HI70300 storage solution or in pH 7.01 buffer solution for several hours. Then follow the calibration procedure:
DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.
Note: The electrode tip should be rinsed with purified water (reverse osmosis, distilled, or deionized) before and after placing in any solution (buffer, storage or sample).
Range | 0.0 to 14.0 pH 0.0 to 50.0°C (32.0 to 122.0°F) |
Resolution | 0.1 pH / 0.1°C / 0.1°F |
Accuracy (@25°C/77°F) | ±0.1 pH / ±0.5°C / ±1.0°F |
Calibration | automatic, one or two-points (pH 4.01, 7.01, 10.01) |
Temperature Compensation | automatic, 0 to 50°C |
Battery Type | CR2032 3V Li-ion (1 pc.) |
Battery Life | approximately 800 hours of continuous use |
Environment | 0 to 50°C (32 to 122°F); RH 100% max |
Dimensions | 160 x 40 x 17 mm (6.3 x 1.6 x 0.7") |
Weight (without battery) | 65 g (2.3 oz.) |
Press the ON/OFF button to turn the meter on. At start-up, all the LCD segments are displayed for 1 second, then the percent indication of the remaining battery life is displayed for another second. The meter then enters the normal measuring mode.
Note: Keeping the ON/OFF button pressed while turning the meter on will display all LCD segments as long as the button is pressed.
Press the CAL button. "CAL" message is displayed.
While in measurement mode, remove the battery cover and press the button inside the battery compartment.
While in measurement mode, remove the battery cover. Press the Setup button located on the side of the battery in the battery compartment. The meter will enter in setup mode. Press the ON/OFF button to move through setup parameters. Press the CAL button to change the settings.
Setup Button
The default settings are: "Set t" measure unit - °C, "AOFF" - 8 min. After the last setting, it will exit Setup.
To select the temperature unit (°C/°F)
To select the measurement unit when "Set t" is displayed press the CAL button to change between °C or °F.
To select the Auto-Off time
To select AOFF TIME unit when "AOFF" is displayed press the CAL button to change between 8 min, 60 min or --- (disabled).
To return to measurement mode
Press ON/OFF button to exit the menu.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Before using Hanna Instruments products, make sure that they are entirely suitable for your specific application and for the environment in which they are used. Any variation introduced by the user to the supplied equipment may degrade the meter's performance. Avoid touching the electrode area. To avoid damages or burns, do not put the meter in microwave oven. For yours and the meter's safety do not use or store the meter in hazardous environments.
Hanna Instruments reserves the right to modify the design, construction, or appearance of its products without advance notice.
Proceed with steps 1 through 3 under single-point calibration using pH 7.01 buffer first. Then follow steps below:
Press CAL button. Meter enters in calibration mode.
Press ON/OFF button and CLR is displayed. The meter will now be at default calibration.
The "CAL" tag will be cleared in measurement mode, indicating the missing calibration, until a new calibration will be performed.
To obtain the highest accuracy for measurements it is important to follow these tips:
General | Soak in Hanna Instruments HI7061 General Cleaning Solution for approximately 1 hour. |
Protein | Soak in Hanna Instruments HI7073 Protein Cleaning Solution for 15 minutes. |
Inorganic | Soak in Hanna Instruments HI7074 Inorganic Cleaning Solution for 15 minutes. |
Oil and grease | Rinse with Hanna Instruments HI7077 Oil and grease Cleaning Solution |
After performing any of the cleaning procedures rinse the electrode thoroughly with distilled water, and soak the electrode in HI70300 Storage Solution for at least 1 hour before taking measurements.
The meter features a low battery indicator. When the battery is running low (under 10%), the battery indicator will blink on the LCD. When the battery is discharged "dEAd bAtt" will be displayed on the LCD for 2 seconds and the meter will turn off.
To change the CR2032 Li-ion battery, turn the battery cover located on the back of the meter counterclockwise to unlock. Remove cover and replace with new battery + side facing up.
Note: Batteries should only be replaced in a safe area using the battery type specified in this instruction manual. Old batteries should be disposed in accordance with local regulations.
pH Buffer Solution
Code | Description |
HI70004P | pH 4.01 buffer solution, 20 mL sachets (25 pcs.) |
HI70007P | pH 7.01 buffer solution, 20 mL sachets (25 pcs.) |
HI70010P | pH 10.01 buffer solution, 20 mL sachets (25 pcs.) |
HI77400P | pH 4.01 & 7.01 buffer solution, 20 mL sachets (10 pcs., 5 ea.) |
HI770710P | pH 10.01 & 7.01 buffer solution, 20 mL sachets (10 pcs., 5 ea.) |
Electrode Cleaning Solution
Code | Description |
HI7061M | General purpose cleaning solution, 230 mL bottle |
Electrode Storage Solution
Code | Description |
HI70300M | Electrode storage solution, 230 mL bottle |
This meter is warranted for a period of one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrode is warranted for a period of six months. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem.
If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.