PC INTERFACE

PC INTERFACE
PC INTERFACE
PC INTERFACE
PC INTERFACE
Data transmission from the instrument to the PC can be done with the HI 92000
Windows
compatible software (optional). HI 92000 also offers graphing and
®
on-line help feature.
Data can be exported to the most popular spreadsheet programs for further analysis.
To connect your instrument to a PC, use an USB cable connector. Make sure
that your instrument is switched off and plug one connector to the instrument
USB socket and the other to the serial or USB port of your PC.
Note: If you are not using Hanna Instruments HI 92000 software, please
see the following instructions.
SENDING COMMANDS FROM PC
It is also possible to remotely control the instrument with any terminal program.
Use an USB cable in according with the model to connect the instrument to a
PC, start the terminal program and set the communication options as follows: 8,
N, 1, no flow control.
COMMAND TYPES
To send a command to the instrument follow the next scheme:
<command prefix> <command > <CR>
where:
<command prefix> is a selectable 16 ASCII character
<command> is the command code.
Note: Either small or capital letters can be used.
SIMPLE COMMANDS
KF1
Is equivalent to pressing functional key 1
KF2
Is equivalent to pressing functional key 2
KF3
Is equivalent to pressing functional key 3
RNG
Is equivalent to pressing RANGE
MOD
Is equivalent to pressing MODE
CAL
Is equivalent to pressing CAL
UPC
Is equivalent to pressing the UP arrow key
DWC
Is equivalent to pressing the DOWN arrow key
Is equivalent to pressing RCL
RCL
SET
Is equivalent to pressing SETUP
CLR
Is equivalent to pressing CLR
OFF
Is equivalent to pressing OFF
CHR xx
Change the instrument range according with the parameter
56