Please read this instruction manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at tech@hannainst.com
Find your local Hanna Instruments Office at www.hannainst.com
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occured during shipment, please contact your local Hanna Instruments Office.
Each HI96722 Ion Selective Meter is supplied complete with:
Note: Save all packing material until you are sure that the instrument works correctly. Any defective item must be returned in its original packing.
For more details about spare parts and accessories see "Accessories".
Range | 0 to 80 mg/L |
Resolution | 1 mg/L |
Accuracy @25°C (77°F) | 1 mg/L ±15% of reading |
Light source | Tungsten lamp |
Light Detector | Silicon Photocell with narrow band interference filter @525 nm |
Method | Adaptation of the turbidimetric method. The reaction between cyanuric acid and the reagent causes a white suspension in the sample. |
Environment | 0 to 50°C (32 to 122°F); max 95% RH non-condensing |
Battery Type | 9V (1 pc.) |
Auto-Shut off | After 10' of non-use in measurement mode; after 1 hour of non-use in calibration mode; with last reading reminder |
Dimensions | 192 x 104 x 69 mm (7.6 x 4.1 x 2.7") |
Weight | 320 g (11.3 oz.) |
Measurement
On Zero Reading:
![]() | Light High: There is too much light to perform a measurement. Please check the preparation of the zero cuvette. |
![]() | Light Low: There is not enough light to perform a measurement. Please check the preparation of the zero cuvette. |
![]() | No Light: The instrument cannot adjust the light level. Please check that the sample does not contain any debris. |
On Sample Reading:
![]() | Inverted cuvettes: The sample and the zero cuvette are inverted. |
![]() | Zero: A zero reading was not taken. Follow the instructions of the measurement procedure for zeroing the meter. |
![]() | Under range: A blinking "0.0" indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvette for reference (zero) and measurement. |
![]() | Over Range: A flashing value of the maximum concentration indicates an over range condition. The concentration of the sample is beyond the programmed range: dilute the sample and re-run the test. |
During Calibration Procedure:
![]() | Standard Low: The standard reading is less than expected. |
![]() | Standard High: The standard reading is higher than expected. |
Other Errors And Warnings:
![]() | Cap error: Appears when external light enters in the analysis cell. Assure that the cuvette cap is present. |
![]() | Cooling lamp: The instrument waits for the lamp to cool down. |
![]() | Battery low: The battery must be replaced soon. |
![]() | Dead battery: This indicates that the battery is dead and must be replaced. Once this indication is displayed, normal operation of the instrument will be interrupted. Change the battery and restart the meter. |
Reagent Sets | |
HI93722-01 | Reagents for 100 tests |
HI93722-03 | Reagents for 300 tests |
Other Accessories | |
HI96722-11 | CAL Check™ Standard Cuvettes (1 set) |
HI740029P | 9V battery (10 pcs.) |
HI731318 | Cloth for wiping cuvettes (4 pcs.) |
HI731331 | Glass cuvetes (4 pcs.) |
HI731335 | Caps for cuvettes (4 pcs.) |
HI93703-50 | Cuvette cleaning solution (230 mL) |
do not validate or calibrate the instrument with standard solutions other than the Hanna Instruments CAL Check™ Standards, otherwise erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at room temperature (18 to 25°C; 64.5 to 77.0°F).
Use the Hanna Instruments CAL Check™ cuvettes (see "Accessories") to validate or calibrate instruments.
Validation
Calibration
Note: It is possible to interrupt the calibration procedure at any time by pressing CAL CHECK or ON/OFF keys.
In GLP mode, the last calibration date can be verified and the factory calibration can be restored.
Factory Calibration Restore
It is possible to delete the calibration and restore factory calibration.
To save the battery, the instrument shuts down after 10 minutes of non-use in measurement mode and after 1 hour of non-use in calibration mode. If a valid measurement was displayed before autoshut off, the value is displayed when the instrument is switched on. The blinking "ZERO" means that a new zero has to be performed.
One fresh battery lasts for around 750 measurements, depending on the light level.
The remaining battery capacity is evaluated at the instrument startup and after each measurement.
The instrument displays a battery indicator with three levels as follows:
If the battery is empty and accurate measurements can't be taken any more, the instrument shows "dEAd bAtt" and turns off.
To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument's battery, follow the steps:
Before using these products, make sure that they are entirely suitable for your specific application and for the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments, this requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument safety do not use or store the instrument in hazardous environments.
HI96722 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
Hanna Instruments reserves the right to modify the design, construction, or appearance of its products without advance notice.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Hanna Instruments HI96722 - Cyanuric Acid ISM Manual