EC EC EC EC EC CALIBRATION
CALIBRATION
CALIBRATION
CALIBRATION
CALIBRATION
It is recommended to calibrate the instrument frequently, especially if high
accuracy is required.
The EC range should be recalibrated:
• Whenever the EC electrode is replaced.
• At least once a week.
• Before USP measurement.
• After testing aggressive chemicals.
• When calibration alarm time out is expired - "CAL DUE" blinks (if feature
is enabled in SETUP).
• If "Outside Cal Range" message blinks during EC measurement (the
measurement range is not covered by current calibration, if feature is
enabled in SETUP).
Note: TDS and Resistivity readings are automatically derived from the EC
reading and no specific calibration is needed.
PROCEDURE
HI 98188 instrument offers a choice of 7 memorized standards (0.00 mS/cm,
84.0 mS/cm, 1.413 mS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and
111.8 mS/cm).
For accurate EC measurements, it is recommended to perform a calibration in
maximum allowed points. However, at least a two-point calibration is suggested
(offset calibration in 0.00 mS and one in any other calibration standard).
The instrument will automatically recognize the standards and skip the stan-
dard used during calibration.
• If a five-point calibration is performed, one of the points must be 0.0 mS (offset).
• Pour small quantities of selected standard solutions into clean beakers. For
accurate calibration use two beakers for each standard solution, the first one
for rinsing the electrode and the second one for calibration.
• Remove the protective cap and rinse the electrode with some of the
standard solution to be used for the first calibration point.
FIVE-POINT CALIBRATION
FIVE-POINT CALIBRATION
FIVE-POINT CALIBRATION
FIVE-POINT CALIBRATION
FIVE-POINT CALIBRATION
• It is better to perform the first point calibration as offset calibration.
• Tap the probe repeatedly to remove any air bubbles that may be trapped
inside the sleeve.
• For offset (zero) calibration, just leave the dry probe in the air.