Hanna Instruments HI 98188 Instruction Manual page 62

Usp compliant ec, tds, nacl, resistivity, temperature meter
Table of Contents

Advertisement

For USP records:
• Record ID (3 chars)
• Stage 1 status (1 char): 0 - not verified; 1 - USP met;
2 - USP not met
• Stage 2 status (1 char): see Stage 1
• Stage 3 status (1 char): see Stage 1
• Stage 1 data (if exist):
• EC reading (8 chars)
• EC unit (1 char)
• temperature reading in ºC
• Stage 2 data (if exists) - same as Stage 1
• Stage 3 data (if exists)
• pH value (5 chars)
•USP factor (3 chars)
•Cell constant (1 char)
Note: "Err3" is sent if record does not exist
"Err4" is sent if the identification range character is not recognized
"Err5" is sent if autolog is in progress
"Err6" is sent if the request range is not available
"Err7" is sent if the instrument is logging
"Err8" is sent if the instrument is not in the measuring mode
"Err9" is sent if the battery low condition is detected and the commu-
nication cannot be done.